Cite this post as:
Scott Weingart, MD FCCM. EMCrit Wee – ETCO2 with EGA?. EMCrit Blog. Published on April 22, 2012. Accessed on April 27th 2024. Available at [https://emcrit.org/emcrit/emcrit-wee-etco2-with-ega/ ].
Financial Disclosures:
Dr. Scott Weingart, Course Director, reports no relevant financial relationships with ineligible companies.
This episode’s speaker(s), (listed above), report no relevant financial relationships with ineligible companies.
CME Review
Original Release: April 22, 2012
Date of Most Recent Review: Jan 1, 2022
Termination Date: Jan 1, 2025
You finished the 'cast,
Now Join EMCrit!
As a member, you can...
- Get CME hours
- Get the On Deeper Reflection Podcast
- Support the show
- Write it off on your taxes or get reimbursed by your department
.
Get the EMCrit Newsletter
If you enjoyed this post, you will almost certainly enjoy our others. Subscribe to our email list to keep informed on all of the Resuscitation and Critical Care goodness.
This Post was by the EMCrit Crew, published 12 years ago. We never spam; we hate spammers! Spammers probably work for the Joint Commission.
Hi Scott Just read that porcine arrest model study on carotid blood flow and three EGA airway devices, King alt, Combitube and LMA, that Dr Yannaopoulos and others were talking about in the CPR blog post a while ago. Interesting results…did seem to indicate worse carotid flow cw with no airway or ETI. Pig airway anatomy is similar but smaller to humans. There was a Korean EMS study that came out in March comparing airway management in OHCA outcomes and LMA did worse than BVM or ETI. I thought it was because the LMA can be tricky to place at… Read more »
It can’t hurt to use the ETCO2 on the end of an LMA but I am also not sure it will be overly helpful given it is a low pressure seal prone to leak (which in theory ETCO2 can help you detect). A large Japanese study that found no difference in survival between OHCA patients who were intubated vs those who had an LMA placed (1). This topic is one I’m not really able to quantify my thoughts clearly on so I’ll step to the side and see what others think. (1) Kajino K, Iwami T, Kitamura T, Daya M,… Read more »
Remember, our ETCO2 devices suck up but a tiny portion of each breath. A leak is no problem unless it is causing sig. effects on the actual ventilation of the lungs.
Hi Minh, it’s Jim! I use the monicker “Jimmy D” to keep google from creating new links on me.
Any EGA/airway device that improves ventilation ease and seal (over mask ventilation) can ironically worsen neurologic recovery following resuscitation because it will unwittingly allows the rescuers to hyperventilate with impunity. The work of Aufderheide demonstrated this here in Milwaukee. The LMA outperformed the BVM in its ability to ventilate in the Korean study, but the outcome was worse because of hyperventilation? What do you think?
This is merely an anecdote, but KingLT usage is very common in OHCA in North Carolina. Additionally, waveform EtCO2 is common among the high-performing systems. I believe if there were an issue with their combined use or accuracy we would have heard about it (from Wake Co/New Hanover Co/MEDIC in Charlotte). Granted, that doesn’t exactly prove anything 🙂
Another reason to carry a bougie, stick it through that EGD, into the trachea and slide your ETT into place.
Regarding the lma neuro issue, I think till research rules out this complication,one might be better off either with the ett or bvm.
Small correction, the comment is not limited to the lma but also to other supraglotyic airways
OK, let me see if I understand what you’re asking. EtCO2 monitoring with SGA’s during CPR(?) Are you concerned with the accuracy in terms of diagnosing adequacy of CPR/resuscitation techniques? Otherwise, it will work to indicate that you have a patent airway of course. Ongoing EtCO2 will allow you to ensure you are ventilating.
the question is: can the termination rule of ETCO2<10 at the 10-20 minute mark and the return of rosc rule (increase >10, stop and check pulse) still be used with EGA.
It should apply if the ventilation is carefully handled–let me explain. If the EGA’s are handled within their performance limits, accurate information should be available through EtCO2 monitoring. EGA seals are as such: LMA Classic design can handle 20 cm H2O peak pressure, Second generation LMA’s (LMA Supreme, Air-Q) can handle a bit more, perhaps 25 cm H2O reliably, perhaps as much as 30 cm H2O, and the double balloon EGA’s probably are in the same range as the second generation LMA devices. If the ventilation is performed in a manner during CPR in which there is not a lot… Read more »
Scott, thanks for extending on this topic. Great discussion here as well. This is a very interesting topic to me. Ever since the last European Resuscitation Council guidelines came out in 2010, we started using EGA (i-gel in our case – not affiliated with them in any way) to perform uninterrupted chest compressions during CPR. Guidelines say only this: “Once a supraglottic airway device has been inserted, attempt to deliver continuous chest compressions, uninterrupted during ventilation. If excessive gas leakage causes inadequate ventilation of the patient’s lungs, chest compressions will have to be interrupted to enable ventilation (using a CV… Read more »
There is a technological solution to the problem of timing ventilations with continuous chest compressions–it’s the Oxylator. With mask, EGA or tracheal tube, it will ventilate on the decompression phase of CPR with a fixed minute ventilation of 10-12 lpm. The Oxylator FR-300 sells for $500 USD here in the states, and there are plenty of European distributors (where it is more popular, by the way). I’ve used the Oxylator in 2 cases of cardiac arrest (one with a Combitube and one with a tracheal tube) and it performed effective ventilations throughout the arrests.
We’ve recently had issues with obtaining ETCO2 values when using an Igel and a PEEP valve. Surprisingly without the PEEP valve there’s ETCO2 readings. Theory is that most of the gas is leaking from around the Igel due to an inadequate seal. Anyone have similar experiences?