
COVID-19 might out-strip the number of mechanical ventilators available to us. This has led to interest in using a single ventilator to support multiple patients. This post will review the theory and evidence regarding this (with the admission that I don’t have experience with this).
basic principles
Bedrock principle: Patient-Ventilator Independence
Normally, we adjust the ventilator so that the ventilator is adapted well to suit an individual patient’s needs. The adage is “fit the ventilator to the patient, don’t fit the patient to the ventilator” – in other words, adjust the ventilator to keep the patient comfortable, rather than over-sedating the patient to tolerate an uncomfortable ventilator mode.
We cannot do this when splitting the ventilator. In fact, any interaction where the patient drives the ventilator is problematic (because this allows one patient to affect another patient’s ventilation). For example, we wouldn’t want one patient’s tachypnea to cause other patients attached to the same ventilator to be hyperventilated.
Thus, a bedrock principle of multiple-patient ventilation is that each patient should have no effect on the ventilation of other patients attached to the ventilator. This is achievable, as described below. The alternative is chaos.
Pressure-cycled ventilation is superior to volume-cycled ventilation
The debate regarding pressure-cycled ventilation versus volume-cycled ventilation is perpetual in critical care. Most units and physicians have some preference, but either strategy works well for most patients. In short, the advantages of each are as follows:
- Volume-cycled ventilation: Advantage is delivery of a guaranteed tidal volume (disadvantage is lack of control over peak pressure).
- Pressure-cycled ventilation: Advantage is guaranteed limitation of the peak pressure (disadvantage is lack of control over tidal volume).
Once we start splitting a ventilator between multiple patients, this debate largely evaporates. Using a volume-cycled mode has numerous, major disadvantages:
- Using a volume-cycled mode with multiple patients provides no control over the tidal volume of any patient, and also provides no control on the maximal airway pressure. This is literally the worst of both worlds.
- A volume-cycle mode will introduce the possibility of deleterious interactions between patients. For example, let’s suppose Patient A’s endotracheal tube gets kinked. This will cause Patient B to receive dangerously large tidal volumes!
- Patients sharing the ventilator must have similar size, similar FiO2 and similar PEEP requirements.
Using a pressure-cycled mode solves these problems:
- With a pressure-cycled mode, we retain control over the maximal airway pressure and the driving pressure. We cannot deliver a guaranteed tidal volume to any patient, but that is no different from having any patient on pressure-cycled ventilation. Ability to control and limit the driving pressure may allow this strategy to be reasonably lung-protective.1
- Deleterious interactions between patients are avoided using a pressure-cycled mode. For example, if Patient A’s endotracheal tube gets kinked in a pressure-cycled mode, then Patient A will receive a reduced tidal volume. However, this will have no impact on Patient B.
- Patients sharing the ventilator don’t have to have a similar size. Larger patients will tend to have a greater absolute compliance, so they will receive larger breaths.
Continuous mandatory ventilation (CMV) is required
Normally, patients are able to trigger the ventilator to deliver a breath. This isn’t possible if a single ventilator is being used to support multiple patients (because, as mentioned earlier, one patient’s tachypnea could cause all patients attached to the ventilator to be hyperventilated).
Therefore, the mode of ventilation which must be used is continuous mandatory ventilation (CMV). What this means is that the ventilator fires at a set rate. The patient has no control over the respiratory rate (i.e. the patient cannot trigger a breath). This is an antiquated mode of ventilation, because it’s generally uncomfortable. However, it’s the only way to achieve patient-ventilator independence.
Modern ventilators may lack a continuous mandatory ventilation mode. However, the same effects might be achieved as follows:
- Increase the ventilator trigger threshold as high as possible, so that it’s impossible for patients to trigger a breath (a.k.a. “lock out” the ventilator).
- If #1 is unsuccessful, respirolytic sedation (using drugs that suppress respiratory drive such as opioids and propofol) could be used to reduce patients’ respiratory drive and prevent them from triggering the ventilator. Paralysis would be used only as a last resort.
Ventilation efficacy will be sub-optimal
Carbon dioxide clearance will not be optimized by a multi-ventilator strategy for a few reasons:
- Tidal volumes will be difficult to track and optimize.
- Y-site connections and tubing increase dead space
It will probably be necessary to accept permissive hypercapnia. For patients with substantial acidosis, IV bicarbonate may be needed to support pH (more on this here). As is usually the case in ARDS, the primary focus is providing lung-protective ventilation, rather than optimal blood gas parameters.
Management of profound hypoxemia: Pressure Control Inverse Ratio Ventilation (PC-IRV)
Patients with COVID-19 appear to be relatively responsive to PEEP. Of course, PEEP is merely one way to increase the mean airway pressure – which is the most important variable affecting lung recruitment. For profound hypoxemia, inverse ratio ventilation may be used to increase the mean airway pressure even further.
Inverse ratio ventilation involves increasing the inspiratory time, so that the patient is spending most of the time in the inspiratory phase (inspiratory time > expiratory time). Inverse ratio ventilation is generally not used because it’s uncomfortable, but in this context patients will be deeply sedated anyway. The overall concept here is similar to APRV – trying to maintain an “open lung” with ongoing application of gentle amounts of pressure (rather than abrupt high-pressure recruitment maneuvers).
basic setup to split a single ventilator
So, what this leaves with is the following:
- Multiple patients attached to a single ventilator. The patients don’t need to be the same size, but ideally they should have roughly similar severity of lung injury (e.g. similar PEEP and FiO2 requirements)(more on achieving this matching below).
- The ventilator is set to pressure-cycled ventilation with a high PEEP (noting that patients with COVID-19 seem to be highly PEEP-responsive) and a low driving pressure (to achieve lung protection). For example, a setting of 30 cm / 18 cm might be reasonable for many patients.
- The ventilator trigger is locked out, to prevent patients from triggering breaths.
- Patients will likely require deep sedation to render them passive on the ventilator (e.g. propofol plus opioids). Paralysis isn’t necessarily required, but it may be necessary in some cases, depending on how sensitive patients are to sedation.
- Ventilation efficacy of each patient can be tracked using an end-tidal CO2 monitor placed in-line with their own endotracheal tube (in a shortage of etCO2 sensors, it might be possible to use a single sensor and rotate it between patients to spot-check the pCO2 of each patient sequentially).
- Permissive hypercapnia will need to be anticipated and managed, as discussed above.
- Viral filters should be used to prevent cross-contamination of pathogens between different patients.

In theory, a single ventilator could be used to support multiple patients (e.g. 2-4 patients, possibly even 6 or 8?). At some point the ventilator may not be powerful enough to support the summed tidal volumes of all the patients.
bigger picture: Five ventilators to provide personalized settings to 20 patients
A major drawback of the above setup is that patients must be matched based on relative severity of lung injury (PEEP and FiO2 requirements). This issue could be overcome as follows.
Imagine that we set up five ventilators:
- Ventilator 1: Mild injury settings (FiO2 50%, PEEP 10 cm, Peak pressure 20 cm)
- Ventilator 2: Moderate injury settings (FiO2 60%, PEEP 14 cm, Peak pressure 26 cm)
- Ventilator 3: High injury setting (FiO2 80%, PEEP 18 cm, Peak pressure 30 cm)
- Ventilator 4: Refractory hypoxemia settings (FiO2 100%, PEEP 22 cm, peak pressure 35 cm).
- Ventilator 5: Salvage settings (FiO2 100%, PEEP 22 cm, peak pressure 35 cm, inverse ratio ventilation with inspiratory time >> expiratory time).
Each of the ventilators could be connected to 1-4 patients. If patients deteriorated, they could be moved to a higher-number ventilator (e.g. from Ventilator #2 to #3). Alternatively, as patients improved, they could be shifted to a lower-number ventilator. This system could allow a handful of ventilators to provide reasonably personalized settings to a large number of patients.
evidence?
This general concept has been demonstrated using animals and lung models.2,3 However, depending on exactly how the model is constructed and what gauge is used to determine “success,” different results can be obtained. For example, Branson et al. demonstrated that multi-lung ventilation cannot be used to deliver precise tidal volumes.4 That’s wholly predictable based on physics, so it doesn’t actually reveal anything. So, if we are using delivery of a fixed tidal volume as a gauge for success, then multi-patient ventilation will fail. However, if we are using delivery of a fixed driving pressure as a gauge for success, then multi-patient ventilation may succeed.
One published report does describe the use of split ventilation in two volunteers (using a facemask interface, rather than intubation). Pressure cycled ventilation was successfully applied with good results.5

- It is almost certainly possible to ventilate several patients with a single ventilator. This probably can be achieved with reasonably lung-protective settings (i.e. low driving pressure). However, the cost of this strategy is loss of control over precise tidal volumes and suboptimal ventilation (with high pCO2).
- A fundamental goal of multi-patient ventilation is to prevent any patient from affecting the other patients. This may be achievable using pressure-cycled ventilation without the ability of any patient to trigger the ventilator.
- Patients would need to be deeply sedated and passive on the ventilator (or paralyzed if necessary).
- Each patient’s ventilatory efficiency could be monitored using end tidal CO2. This would be required as a surrogate for tidal volume or minute ventilation (which will not be measurable).
- By using a small number of ventilators with a range of different settings, a large group of patients could be supported with fairly personalized settings.

going further
- COVID IBCC chapter here (with additional links to other COVID resources at the bottom)
- Columbia Presbyterian protocol for splitting ventilators here.
- More advanced and granular exploration about how to hook everything up here.
references
- 1.Amato M, Meade M, Slutsky A, et al. Driving pressure and survival in the acute respiratory distress syndrome. N Engl J Med. 2015;372(8):747-755. doi:10.1056/NEJMsa1410639
- 2.Neyman G, Irvin C. A single ventilator for multiple simulated patients to meet disaster surge. Acad Emerg Med. 2006;13(11):1246-1249. doi:10.1197/j.aem.2006.05.009
- 3.Paladino L, Silverberg M, Charchaflieh J, et al. Increasing ventilator surge capacity in disasters: ventilation of four adult-human-sized sheep on a single ventilator with a modified circuit. Resuscitation. 2008;77(1):121-126. doi:10.1016/j.resuscitation.2007.10.016
- 4.Branson R, Blakeman T, Robinson B, Johannigman J. Use of a single ventilator to support 4 patients: laboratory evaluation of a limited concept. Respir Care. 2012;57(3):399-403. doi:10.4187/respcare.01236
- 5.Smith R, Brown J. Simultaneous ventilation of two healthy subjects with a single ventilator. Resuscitation. 2009;80(9):1087. doi:10.1016/j.resuscitation.2009.05.018
- PulmCrit blogitorial: Why I don't believe the AMIKINHAL trial - October 26, 2023
- PulmCrit Blogitorial – New IDSA/SCCM guidelines on fever evaluation in ICU - October 24, 2023
- PulmCrit Blogitorial – Why the new brain death guidelines are unacceptable - October 13, 2023
Have you seen this, Josh?
https://www.youtube.com/watch?v=uClq978oohY&feature=youtu.be
Petros
Hi Dr. Kopterides! Nice to see the ICU closest to me thinking ahead 🙂 I got in contact with Dr. Menes and hopefully he will do an online lecture soon to talk about his experience venting multiple patients. We do have two ID lecturers coming up if anyone at Excela is interested: https://www.health4theworld.org/covid/
What happened in Vegas….. https://epmonthly.com/article/not-heroes-wear-capes-one-las-vegas-ed-saved-hundreds-lives-worst-mass-shooting-u-s-history/
The MD used a daisychain setup – one tube out (the stem), T’d into a circle (central flower). 8 branches coming off the circle, beds arranged like petals of the flower.
So you would manipulate driving pressure by adjusting PEEP based on plateau pressures ?
exactly.
driving pressure ~ [ (peak pressure) – peep ]
in a pressure cycled mode you could control this and maintain within a safe range.
I currently travel to work at a rural CCA that primarily transports out to a nearby trauma center, but that may not be an option for long, 2 vents and no real dedicated ICU in house. Hopefully I won’t ever have to use this, but I could see it being especially useful to larger institutions, as they will have more vents and more of an ability to group patients effectively. Thoughts on the article (with zero experience doing this) Modifying circuits for off label use may change how well the vent functions, beyond just the question of adequate pressure/flow for… Read more »
can the ventilator hook to pressurized air chamber “lung”, one per person that then release into the tube upon individual requirements?
Do you know of any techniques on how to perform split ventilation method when using single limb King Circuits?
I don’t know how this could be done, sorry
Hello, It was a good experience while reading your blogs. Thanks for sharing with us.
You still have the same issue with pressure ventilation? On a dual system gas is going to travel to the path of least resistance regardless of Vol vs. Press ventilation. If the pressure setting is reached faster in subject B with less compliant lungs, the system will cycle into expir too soon to effectively ventilate both subjects.
you would have to set a high enough inspiratory time (i-time) to make sure that pressure equilibrated throughout all the patients (ideally end-inspiratory flow would drop to zero prior to the termination of the breath). these patients do well with high i-times anyway, because that increases the mean airway pressure and is good for recruitment.
It would seem possibly safer and simpler to hook up a patient post intubation to a bag and peep valve with a blender to dial in the oxygen concentration . Should be able to monitor with a pulse ox. Could train nonmedical people to bag at a certain rate. Would be unlimited number of patients who could be treated even in non -ICU bed .
This is my thought as well as I’ve seen this strategy grow on commentary popularity. I can not imagine moving ARDS patients to and from “ventilator pods” as they transition in and out of compliance changes. Also assuming we would rely entirely on ABG and possibly mixed venous results for these transitions as there will be zero reliable information coming from the ventilator as far monitoring these compliance changes. The whole thing is more frightening than enlisted family members or volunteers and equipping them to manually bag for periods of time.
you have no precise control over peak pressure, plateau pressure, tidal volume, or really anything else with manual bagging. doesn’t seem like a great option. also whoever is doing the bagging will burn through lots of PPE.
Your explanation adds considerable detail to this Youtube video https://www.youtube.com/watch?v=uKtEI-fidD0 I can confirm that people are either using this approach or looking at it. Governor Cuomo said something about supporting 2 patients per ventilator. Your approach provides graded support, which is flexible by design. Please get in touch with colleagues in New York, if you have not already. Sorry for the push, but who knows if they have had any time to review your notes. Thank you for your patience in reading this post. I wish you all the best. Kent keywords: multiple patients ventilator multiplexing graded treatment (that’s for… Read more »
Hi–I have been working with Professors Chen Sun and Jiaxing Huang in the McCormick School of Engineering at Northwestern University. They are in the process of 3D printing manifolds that might make this process easier in a disaster situation. We are in the process of testing and doing dry runs. It is our hope that these devices can be sanitized well enough to be considered safe to use with multiple patients with COVID19 respiratory failure.
that’s fantastic!! great work !!
Hello.
Can we set a different volume and different Fio2 on one ventilator???
Hi Michael, have you experimented with the models I published on Prusa Printers? I’d love your feedback. The 4 way splitter is not pictured, but is available for download. https://www.prusaprinters.org/prints/25808-3d-printed-circuit-splitter-and-flow-restriction-d
Hi Michael,
Great idea! We’ve just launched a platform where we collect 3D printable parts in a digital inventory and also connect hospitals with 3D printing suppliers in the nearest area.
You’re welcome to add your parts to our inventory! https://www.3yourmind.com/news/3d-part-order-management-corona-response
Great info on multiple patient setups. Seems like it would be possible to design connectors used in such setups to include check valves and adjustable pressure relief valves in order to allow as least some customization of individual patient settings. Similarly, it would also seem possible to increase FiO2 levels on individual lines with a variable valve based on inputs from an oximeter or CO2 sensor. I’m an engineer, not a medical provider. It’s clear to me that producing these components (and the interconnect tubing) in mass is much more realistic than producing tens of thousands of new ventilators in… Read more »
Sorry…on prior post, I meant “Defense Production Act”, not the “War Powers Act”.
Some help from FDA relaxing regulations/enforcement during the crisis at this link: https://www.medtechdive.com/news/coronavirus-fda-relaxes-rules-on-ventilator-manufacturers/574648/
I think they need to go further. Force manufacturers to provide model specific-guidance that will increase the odds of success when multiplexing their equipment…and indemnify them of inherent risks. Without this, their only response will continue to be that multiplexing is not advised.
John, Like you, I am an engineer not a medical professional, but I had a few thoughts about sharing ventilators between patients with mismatched lung compliance. Flow restrictors might help match otherwise mismatched patients, but a regulator/pressure drop valve on line line would likely be better. If neither is available, I wondered if it might be possible to modify effective lung compliance by actually loading a patient’s chest with elastic straps, weighted blankets, weight plates or water bags. Pressures seem to be in the range of 5-20 cm of water. My chest moves by ~ 1cm when I take a… Read more »
With ventilators in short supply, NewYork-Presbyterian Hospital, one of the city’s largest systems, has begun using one machine to help multiple patients at a time, a virtually unheard-of move, a spokeswoman said. -NY TIMES 3/25/2020
THINKING OUTSIDE THE BOX The tubing circuitry complexity to a number of patients with not only likely different compliance at baseline but also changes with time, risks hyperventilation of one patient and simultaneous hypoventilation of another. This variability is amplified with more patients sharing the same source of positive pressure. There may be a better way. Borrowing from hydrodynamics of public water works, a single main supply is tapped into with each individual house, the flow metered by faucet valves. We could have a high capacity positive pressure source- a pump with peak pressure of say, 45 cm H20, Also… Read more »
To follow on Dr Petersen’s comment, consider the idea of a “modulating module” on each splitter line consisting of a manual pinch valve and a differential pressure sensor (before and after valve) which enables specific pressure support adjustment and also an estimate of tidal volume (thru calibration of pinch valve characteristic vs delta P). More complex than just an arrangement of splitter connectors – but much more functional I think.
Read with interest the post on using a ventilator to support multiple patients. How are patients effectively weaned off?
I propose a superior method would be to utilize APRV/Bilevel to allow each patient to breath spontaneously independently of each other. In this mode the vent will simply cycle between P(high) and P(low) at set time intervals regardless of any patients individual respiratory rate. By design, none of the patients will be able to inhale during the P(low) cycle, but we know from experience with APRV/Bilevel that this mode is well tolerated in conscious patients. This would theoretically eliminate the need for heavy sedation and/or paralytics to prevent spontaneous respirations. Any thoughts on this suggestion would be appreciated as I… Read more »
https://www.aarc.org/wp-content/uploads/2020/03/032620-COVID-19-press-release.pdf?utm_source=informz&utm_medium=email&utm_campaign=aarc-protech&_zs=675RO1&_zl=f3if5
While this method may have proven useful in trauma patients with healthier lungs, it is not ideal for this disease process that specifically attacks the lungs. You will be better off to chose the patient to most likely have the better outcome, and ventilate them optimally. Above is a statement from the AARC.
New experimental data out today: https://www.mn.uio.no/fysikk/english/research/projects/bioimpedance/splitting_ventilator.pdf
I am retired and practiced as a Respiratory Therapist as well as being the Director of Respiratory Care for 20 years. I certainly understand the need for more ventilators. However, ventilators do not operate by themselves, and Nurses do not always understand the function of ventilators. Remember, too many fingers in the pie. Those are some very long sections of tubing. Would the long tubing not create more resistance? Tubing/connectors have a habit of becoming disconnected. I would not like to be the patient who has been paralyzed to control my breathing and the tubing becomes disconnected (which I have… Read more »
Thanks for putting this together. This should serve as a great repository if the need arises. Naturally this would be a (very) last resort, but it helps to think about it now. Some additional considerations: Patients may need to be split not just by disease severity but by COVID phenotype as well (Type L would need less PEEP than Type H). Patients may move down to pods with less support or back up pods if they deteriorate. The easiest pods to manage, and perhaps the only pods needed may be the high sedation, high FiO2, +/- paralysis pods, as there… Read more »
Dr. Farkas,
I’m not sure if you have come across the PReVents protocol out of Yale which uses a pressure-controlled mode with modified in-line pressure-gated valves to allow for individualized PIP and PEEP. They seem to address a lot of the issues with the multidisciplinary statement against ventilator splitting.
Here’s the link to the pre-print: https://www.medrxiv.org/content/10.1101/2020.04.03.20052217v1.article-metrics
I’m one of the authors. I think we used a lot of the ideas that other people have brought up in the comments.
need help, can we use ventilator with multiple tubing, in acute oxygen shortage??
I truly appreciate that you are doing great work by providing valuable information through your blogs to the people
Mahindra 245
” Universities in Russia,
Russia has been always a preferred destination for Indian students for persueing MBBS abroad.
Best Resort In Lonavala With Pool: 2022 | Sereniity Resort
Sereniity Resort is situated at a prime location in Tungarli, Lonavala.
I really appreciate your work. I am very impressed by this post. I would like to suggest you that please keep sharing such type of info. Thanks.. eicher tractor 242
Good.. I am very impressed by this post. I would like to suggest you that please keep sharing such type of info. Thank you… kubota mu4501
इन्व्हेस्टमेंट साठी सर्वात उत्तम प्रोजेक्ट चाकण MIDC जवळ
Really Instructive Content, Keep Sharing
I really liked your post and you keep posting like this. learned something from your post
<a href=”https://www.trucksbuses.com/scv/pickups-and-mini-trucks/tata-ace-gold-diesel-plus“>Tata ace gold diesel plus</a>
Great post! I really enjoyed reading your article and found it very informative. I especially liked your point about the topic. Keep up the excellent work, and I look forward to reading more from you!
Tata Magic Express
Very useful and knowledgeable blog. I learn so many things from here. Thanks for sharing this blog with us.
Tata Yodha Price
I enjoy to read each word in the article.
Osm Three Wheeler
Very Good Blog . The way of presenting is superb
Tata Tipper
Great job on writing such an insightful and informative article! Your attention to detail and thorough research really shines through in your writing. I appreciate how you presented the information in a clear and organized manner, making it easy for readers to understand and follow along.
Mahindra Trucks
Hello Everyone
Impressive Post, I appreciate your Content on your website.
Electric 3 Wheelers
The Tata Ace Diesel is also known for its durability and reliability, making it a preferred choice for small businesses and entrepreneurs. With its impressive mileage and low maintenance cost, the Tata Ace Diesel is an excellent investment for those looking for a cost-effective mode of transportation.
Teak Furniture Malaysia operates as a specialist importer of best quality <a href=”https://teakfurniture.my/”> wooden furniture malaysia </a> solid wood furniture, indoor furniture, Garden Furniture and outdoor furniture . Visit Our Showroom Kuala Lumpur KL and Shah Alam Selangor Malaysia.
Teak Furniture Malaysia operates as a specialist importer of best quality
https://teakfurniture.my/