Untitled 0
Pulse Oximetry Part I by Stephen Colucciello, MD
Sunday, May 9, 2010
Is it necessary to transport arterial samples on ice?
Basically, the answer is “NO.” In the first study, the authors conclude that
blood samples analyzed within 20 minutes of arterial puncture do not require
storage on crushed ice (with the exception of those obtained in patients with
very high white cell counts). In the second paper, the authors suggest that
arterial blood samples do not require storage in ice if blood gas analysis is to
be performed within 30 minutes.
17. IS IT NECESSARY TO TRANSPORT ARTERIAL SAMPLES ON ICE FOR pH AND GAS
ANALYSIS? Nanji, A.A., et al, Can Anaesth Soc J 31(5):568, September 1984
Studies comparing the effect of storage on ice and storage at room temperature
on the pH, PaO2, and PaCO2 of arterial blood samples have yielded conflicting
results. This study, from the University of British Columbia, compared these
parameters in duplicate arterial blood samples obtained from 21 ICU patients.
Blood samples of 5-6ml, obtained with syringes having a heparin volume of
approximately 0.1 ml, were stored at room temperature or on crushed ice (0øC).
Serial determination of pH, PaO2, and PaCO2 was performed up to 60 minutes after
the samples were obtained. All of the patients had normal hemoglobin values, and
none had an elevated white cell count. The pH values ranged between 7.17 and
7.64. The ranges of the PaO2 and PaCO2 were 54-230mm Hg and 24-77mm Hg,
respectively. In both sets of blood samples, increasing storage time was
associated with decreases in the pH and PaO2, and an increase in the PaCO2.
However, the difference in the magnitude of the changes between the two storage
methods was not considered to be clinically significant until 20 minutes after
the samples were obtained. in samples stored on ice for 20 minutes, the range of
pH changes was 0 to minus 0.02 units; in samples stored at room temperature for
20 minutes, the range was 0 to minus 0.03 units. The range of the changes in
PaO2 and PaCO2 after 20 minutes of storage on ice or at room temperature were
similar. The authors conclude that blood samples to be analyzed within 20
minutes of arterial puncture do not require storage on crushed ice, with the
exception of blood obtained from patients with very high white cell counts,
which is associated with an excessive rate of oxygen metabolism and carbon
dioxide production. 6 references 3/85-#23
18. STABILITY OF BLOOD GASES IN ICE AND AT ROOM TEMPERATURE Liss, H.P., et al,
Chest 103(4):1120, April 1993 BACKGROUND: The reported temperature-dependent
changes in arterial blood gas composition that may occur during the interval
between blood sampling and analysis have prompted recommendations that blood
samples be kept in ice prior to arterial blood gas measurement. METHODS: The
authors, from Wright State University School of Medicine in Dayton, OH, compared
changes in blood gas composition in samples stored in ice or at room
temperature. One hundred nineteen arterial blood samples were analyzed
immediately or were stored in ice prior to baseline analysis. After this initial
analysis, 60 samples were stored in ice and 59 were stored at room temperature,
and further blood gas analysis was performed after 15 and 30 minutes of storage.
RESULTS: In both groups of samples, the PaO2 increased significantly after 15
minutes of storage and exhibited a further significant increase after 30
minutes, and the PaCO2 and pH decreased significantly after 15 minutes. In the
sample stored at room temperature, but not those stored in ice, the pH exhibited
a further significant decrease after 30 minutes. The mean PaO2 increased from
79.7mm Hg at baseline to 82.7mm Hg at 30 minutes in samples stored in ice, and
from 78.2mm Hg to 80.6mm Hg in samples stored at room temperature. The PaCO2
decreased from 38.5mm Hg to 38.0mm Hg and from 37.9mm Hg to 37.4mm Hg in the ice
and room temperature groups, respectively, and the pH decreased from 7.43 to
7.42 and from 7.41 to 7.40, respectively. CONCLUSIONS: The authors suggest that
arterial blood samples do not require storage in ice if blood gas analysis is to
be performed within 30 minutes. 16 references 08/93 - #34
What about doing venous gases instead of arterial gases in the setting of
metabolic acidosis and for screening of hypercapnia?
In the first paper, the authors suggest that patients with nearly normal venous
blood gases are unlikely to have severe acid-base disturbances, while extremely
abnormal venous blood gases reflect comparable arterial blood gas values.
However, they note that the predicted arterial pH may differ by more than 0.11
units in 5% of cases, the predicted bicarbonate may differ by more than
4.96mEq/l in 5%, and the predicted pCO2 may differ by more than 14.18 torr in
5%.
In the second paper examining acutely ill ED patients, there was a high degree
of correlation and agreement between venous and arterial pH.
In the third paper, there was excellent correlation between arterial and venous
blood gas samples. However, the emergency physicians who were surveyed felt that
the actual differences were too large to allow the routine substitution of
venous for arterial gases.
In the fourth paper, the authors found that among patients with acute
respiratory illness, venous pH correlates well with arterial pH. However, levels
of agreement between the arterial and venous pCO2 were less optimal in many
individuals. The authors suggest that a combination of venous blood gas analysis
and pulse oximetry can be a useful alternative to routine arterial blood gas
analysis in such patients. As a screen for hypercarbia, a venous pCO2 cut-off of
45mm Hg had a sensitivity of 100% and a specificity of 57%.
19. THE USEFULNESS OF PERIPHERAL VENOUS BLOOD IN ESTIMATING ACID-BASE STATUS IN
ACUTELY ILL PATIENTS Gennis, P.R., et al, Ann Emerg Med 14(9):845, September
1985 The authors analyzed the relationship between acid-base parameters of
paired arterial and venous blood samples in 171 acutely ill adults and twelve
cardiac arrest patients treated at the Bronx Municipal Hospital Center in New
York. Significant alkalemia was present in 28 of the nonarrest patients and
significant acidemia in 20. In the nonarrest patients, the mean venous pH was
0.056+/-0.056 units lower than the mean arterial pH, the mean venous pCO2 was
7.38+/-7.51 torr higher, and the mean bicarbonate was 1.21+/-2.55mEq/l higher
than the arterial level. In the cardiac arrest patients, the differences were
0.086+/-0.10 units, 6.58+/- 19.98 torr, and 3.74+/-14.05mEq/L, respectively.
Equations were developed that allowed relatively accurate prediction of 68% of
arterial pH, pCO2, and bicarbonate values in nonarrest patients based on
determination of these parameters in venous blood. However, the authors point
out that the predicted arterial pH may differ by more than 0.11 units in 5% of
cases, the predicted bicarbonate may differ by more than 4.96mEq/l in 5%, and
the predicted pCO2 may differ by more than 14.18 torr in 5%. The authors
conclude that their findings do not support the routine utilization of venous
blood specimens for assessment of acid-base status, but indicate that
inadvertent peripheral venous blood sampling may prove to be useful in the
management of acid-base problems in acutely ill patients. They suggest that
patients with nearly normal venous blood gases are unlikely to have severe
acid-base disturbances, while extremely abnormal venous blood gases reflect
comparable arterial blood gas abnormalities. 11 references 1/86-#23
20. VENOUS pH CAN SAFELY REPLACE ARTERIAL pH IN THE INITIAL EVALUATION OF
PATIENTS IN THE EMERGENCY DEPARTMENT Kelly A.M., et al, Emerg Med J 18:340,
September 2001 BACKGROUND: Although several small studies have reported that the
venous pH is a reliable substitute for the arterial pH, most patients requiring
evaluation of acid-base status undergo arterial sampling. Arterial puncture is
painful and carries a risk of complications for the patient, as well as
potential needlestick injury for the healthcare provider. METHODS: This
prospective Australian study compared venous and arterial pH values in ED
patients considered by their managing physician to require arterial blood gas
analysis due to an acute respiratory condition (196) or suspected metabolic
derangement (50). RESULTS: The mean arterial pH in these patients was 7.38, with
values ranging from 7.05 to 7.61. Alkalosis was present in 17% of the patients,
and acidosis in 27%. The correlation between the venous and arterial pH was very
close (r=0.92), and the average difference between the two values was only -0.04
units (range, -0.16 to +0.06). Bias plot evaluation also demonstrated good
agreement between arterial and venous pH, with the 95% limits of agreement being
-0.11 to +0.04 units. CONCLUSIONS: In these acutely ill ED patients, there was a
high degree of correlation and agreement between venous and arterial pH. 10
references 3/02 - #20
21. CAN PERIPHERAL VENOUS BLOOD GASES REPLACE ARTERIAL BLOOD GASES IN EMERGENCY
DEPARTMENT PATIENTS? Rang, L.C.F., et al, Can J Emerg Med 4(1):7, January 2002
METHODS: In this prospective study, from Queens University and the University of
Ottawa in Canada, arterial and venous blood samples were drawn from 218 ED
patients requiring blood gas analysis to determine if the venous pH, pCO2 and
bicarbonate are predictive of arterial values. Forty-five board-certified
emergency physicians were surveyed to determine what values they would consider
to represent clinically important differences between venous and arterial pH,
pCO2 and HCO3. RESULTS: There was excellent correlation between venous and
arterial pH, pCO2 and HCO3. The mean differences between venous and arterial
values were small (0.036 units for pH, 6mm Hg for pCO2 and 1.5mEq/l for HCO3).
The 26 emergency physicians responding to the survey felt that about 0.05 units
for pH, 6.6mm Hg for pCO2 and 1.5mEq/l for HCO3 were clinically significant
differences. Sixty-six percent of the pH values, 51% of the pCO2 values and 87%
of the HCO3 values fell within these "acceptable" ranges. CONCLUSIONS: There was
excellent correlation between arterial and venous blood gas samples in these ED
patients. However, the emergency physicians who were surveyed felt that the
actual differences were too large to allow the substitution of venous for
arterial gases. The authors suggest that venous values might be valid if used
with a correction factor, or to follow trends. 12 references 6/02 - #20
22. VENOUS pCO2 AND pH CAN BE USED TO SCREEN FOR SIGNIFICANT HYPERCARBIA IN
EMERGENCY PATIENTS WITH ACUTE RESPIRATORY DISEASE Kelly A.M., et al, J Emerg Med
22(1):15, 2002 BACKGROUND: Several small studies have suggested that arterial pH
and pCO2 can be estimated from venous blood, but most of these were not
conducted in patients with acute respiratory illness. METHODS: This Australian
study compared arterial and peripheral venous pH and pCO2 in paired blood
samples obtained from 196 ED patients with acute respiratory signs and symptoms.
Significant hypercarbia was defined as an arterial pCO2 of 50mm Hg or higher.
RESULTS: Significant hypercarbia was identified in 29% of the patients. Arterial
pH values ranged between 7.15 and 7.61, and arterial pCO2 ranged between
50-147mm Hg. There was very good agreement between arterial and venous pH, with
the latter averaging 0.03 units lower than the former (95% limits of agreement
[minus] 0.10 to [plus] 0.04). On average, the venous pCO2 was 5.8mm Hg higher
than the arterial pCO2 (95% limits [minus] 8.8mm Hg to [plus] 20.5mm Hg). As a
screen for hypercarbia, a venous pCO2 cut-off of 45mm Hg had a sensitivity of
100% and a specificity of 57%. CONCLUSIONS: Among patients with acute
respiratory illness, venous pH correlates well with arterial pH. Levels of
agreement between the arterial and venous pCO2 were less optimal in many
individuals. The authors suggest that a combination of venous blood gas analysis
and pulse oximetry can be a useful alternative to routine arterial blood gas
analysis in such patients. 14 references 5/02 - #37
How well do venous gases correlate with arterial blood gases in DKA?
In the first study, the author suggests that there is reasonable evidence that
the venous and arterial pH are clinically interchangeable in hemodynamically
stable DKA patients without respiratory failure. However, the study consisted of
only 21 patients.
In the second paper, the authors, from the University of Pennsylvania, reviewed
the available literature to determine the clinical utility of ABG measurement in
the initial management of DKA. The results suggest that venous pH reliably
reflects arterial pH, that differences between the two are not likely to be
clinically significant, and that it seems reasonable to substitute venous blood
sampling for arterial blood sampling in patients with suspected DKA.
In the third study, the authors suggest that venous blood may be used as an
alternative in patients with uremia or DKA, and that the addition or subtraction
of the mean differences noted in their study will provide an accurate
approximation of arterial pH and bicarbonate levels.
The final study in this section, a convenience sample of 38 patients with 44
episodes of DKA, suggests that peripheral venous blood gases may be a reliable
substitute for arterial blood gases in the initial assessment of patients with
suspected DKA.
23. THE CASE FOR VENOUS RATHER THAN ARTERIAL BLOOD GASES IN DIABETIC
KETOACIDOSIS Kelly, A.M., Emerg Med Australasia 18(1):64, February 2006
BACKGROUND: Although arterial blood gas sampling continues to be recommended for
the assessment of acid-base status in patients with diabetic ketoacidosis (DKA),
there is some evidence to suggest that pH and, possibly, bicarbonate can be
accurately estimated in venous blood samples. METHODS: This Australian author
summarized the evidence from six studies (1,022 patients) evaluating the
agreement between arterial and venous pH, four of which (784 patients) also
evaluated the agreement between arterial and venous bicarbonate. RESULTS: Three
of the six studies assessing pH values included patients with DKA. In these
studies the weighted average difference between arterial and venous pH was 0.02
units (arterial higher than venous) (95% limits of agreement -0.009 to +0.021
units). In the other three studies that included patients with mixed respiratory
and metabolic illness and respiratory failure, the weighted average difference
was 0.037 units (95% limits of agreement -0.11 to +0.04 units). Only one of the
four studies of agreement in bicarbonate values specified a DKA subgroup (21
patients). The weighted average difference was -1.88mEq/L in the DKA subgroup
and -0.99mEq/L in the remaining patients (95% limits of agreement -2.73 to
+5.13mEq/L). CONCLUSIONS: The author suggests that there is reasonable evidence
that the venous and arterial pH are clinically interchangeable in
hemodynamically stable DKA patients without respiratory failure. Results
regarding agreement between arterial and venous bicarbonate in DKA patients
should be viewed with caution, as they are based on findings in only 21
patients. 15 references 7/06 - #11
24. ARTERIAL BLOOD GAS ANALYSIS: ARE ITS VALUES NEEDED FOR THE MANAGEMENT OF
DIABETIC KETOACIDOSIS? Kreshak, A., et al, Ann Emerg Med 45(5):550, May 2005
BACKGROUND: The American Diabetes Association advises an initial arterial blood
gas (ABG) in patients with suspected diabetic ketoacidosis (DKA) to determine
the severity of acidemia, but also recommends serial measurement of venous pH to
monitor the effects of treatment. METHODS: The authors, from the University of
Pennsylvania, reviewed the available literature to determine the clinical
utility of ABG measurement in the initial management of DKA and whether the
venous pH reliably reflects arterial pH as a measure of acid-base status. Four
relevant papers were identified. RESULTS: In a study of 200 patients with
suspected DKA, the ABG influenced the final diagnosis in 1% of cases, altered
treatment in 3.5% (five of these seven involved a change in the route of insulin
administration based on pH) and changed disposition in 1%. Venous pH closely
reflected arterial pH, typically being 0.015 points lower. The second study
compared arterial and venous blood gases in 100 patients with uremia, 21 with
DKA and 31 normal controls, and found that venous pH was typically 0.05 points
lower than arterial pH. The third study compared pretreatment venous and
arterial pH in 44 patients with suspected DKA, and reported a close correlation
between the two values and an average difference of 0.03. The fourth study
compared arterial and finger capillary blood gas results in 20 patients with
DKA, and reported a close correlation between these values (mean difference,
0.03). CONCLUSIONS: The results of this review suggest that the venous pH
reliably reflects arterial pH, that differences between the two are not likely
to be clinically significant, and that it seems reasonable to substitute venous
blood sampling for arterial blood sampling in patients with suspected DKA. 5
references 9/05 - #22
25. COMPARISON OF BLOOD GAS AND ACID-BASE MEASUREMENTS IN ARTERIAL AND VENOUS
BLOOD SAMPLES IN PATIENTS WITH UREMIC ACIDOSIS AND DIABETIC KETOACIDOSIS IN THE
EMERGENCY ROOM Gokel, Y., et al, Am J Nephrol 20(4):319, July-August 2000
METHODS: The authors of this Turkish study measured both bicarbonate
concentrations and pH in simultaneously obtained arterial and venous blood
samples in 100 uremic, acidotic patients with end-stage renal failure, 21
patients with DKA, and 31 healthy controls. RESULTS: Among both the uremic and
the DKA patients, there was almost perfect correlation between venous and
arterial measurements of both pH and bicarbonate levels (r=0.98-0.99). The
venous pH was consistently lower than the arterial pH, by a mean of 0.04-0.05
(range, 0.01-0.10), while bicarbonate levels were higher in venous than arterial
samples, by a mean of just under 2mmol/L (range 0.6-2.8mmol/L). These
relationships were very similar for both groups of patients. In the healthy
controls, on the other hand, the correlation between arterial and venous pH and
bicarbonate levels was far less strong (r=0.58). CONCLUSIONS: Arterial puncture
for the measurement of acid-base status may be associated with complications.
The authors suggest that venous blood may be used as an alternative in patients
with uremia or DKA, and that the addition or subtraction of the mean differences
noted in their study will provide an accurate approximation of arterial pH and
bicarbonate levels. 19 references 2/01 - #25
26. COMPARISON OF ARTERIAL AND VENOUS BLOOD GAS VALUES IN THE INITIAL EMERGENCY
DEPARTMENT EVALUATION OF PATIENTS WITH DIABETIC KETOACIDOSIS Brandenburg, M.A.,
et al, Ann Emerg Med 31(4):459, April 1998
BACKGROUND: Patients with diabetic ketoacidosis (DKA) generally undergo
venipuncture at the initial evaluation for determination of blood chemistry, as
well as arterial puncture for blood gas measurement. Studies in patients without
DKA suggest a close correlation between venous and arterial blood gases.
METHODS: The authors, from the University of Oklahoma Health Sciences Center,
compared venous and arterial blood gas values in a convenience sample of 38
patients with 44 episodes of DKA (defined as an arterial pH below 7.35 or a
serum CO2 below 20mmol/L, serum glucose above 250mg/dL, and positive serum
ketones). RESULTS: Close correlation was observed between the pH of arterial and
venous blood (mean difference, 0.03, range 0-0.11, r=0.97), as well as between
arterial and venous HCO3 (r=0.95), and between arterial HCO3 and venous CO2
(r=0.90). The venous pH differed from the arterial pH by more than 0.10 in only
one of the 44 episodes (in which the difference was only 0.11 and was not
believed to be associated with diagnostic or therapeutic significance).
CONCLUSIONS: Findings in this small convenience sample require validation, but
suggest that peripheral venous blood gases may be a reliable substitute for
arterial blood gases in the initial assessment of patients with suspected DKA.
39 references 8/98 - #19
In the setting of cardiac arrest, can arterial gases be expected to accurately
reflect tissue oxygenation?
The first study, from the Chicago Medical School, compared arterial and venous
blood gases in 17 episodes of cardiac arrest induced in twelve pigs. They
concluded that arterial blood gas analysis does not reflect the acid-base status
of systemic tissue during CPR.
The second paper is a clinical study, from the University of Health/Science
Chicago Medical School, which compared arterial and mixed venous blood gas
parameters during cardiopulmonary resuscitation in 16 patients. The data suggest
that respiratory acidosis occurs during cardiopulmonary resuscitation, which may
be reflected in mixed venous but not in arterial blood gases. The findings
indicate that acid-base management of cardiopulmonary arrest based on ABGs may
be inappropriate.
In the final paper of this section, the authors analyzed arterial and venous
blood gas changes during CPR in 28 pigs. They found evidence of respiratory
alkalosis in arterial blood, reflecting a decrease in pulmonary blood flow, and
an acidosis in mixed venous blood, reflecting accumulation of CO2 proximal to
the alveolar capillary bed. This mixed venous acidosis may more accurately
represent systemic acid-base status.
27. ARTERIOVENOUS CARBON DIOXIDE AND pH GRADIENTS DURING CARDIAC ARREST
Grundler, W., et al, Circulation 74(5):1071, November 1986
The American Heart Association has recommended that arterial blood gases be
employed to determine the need for bicarbonate therapy during resuscitation from
cardiac arrest. This study, from the Chicago Medical School, compared arterial
and venous blood gases in 17 episodes of cardiac arrest induced in twelve pigs.
The animals were subjected to periods of ventricular fibrillation followed by
countershock and mechanical chest compression after the onset of EMD, which
occurred in each case. Serial analysis of arterial and venous blood gases
demonstrated a decrease in arterial pCO2, from 37.4mm Hg at baseline to 20.1mm
Hg two minutes after the onset of EMD, and an increase in pH from 7.46 to 7.54.
However, venous pCO2 was observed to increase from 45.2mm Hg to 54.2mm Hg, and
venous pH decreased from 7.41 to 7.31. This widening of the normal venoarterial
pCO2 and pH gradients was related to decreased pulmonary clearance of CO2 due to
reduced pulmonary blood flow. The venoarterial pCO2 and pH gradients normalized
in animals that were successfully resuscitated. The venous acidemia observed in
the present study is considered to be a reflection of tissue acid-base
conditions. Similar findings have also been reported in humans undergoing CPR.
It is concluded that arterial blood gas analysis does not appear to reflect the
acid-base status of systemic tissue during CPR. The authors also suggest that,
on the basis of these findings, bicarbonate administration, which increases the
CO2 load, would not appear to be routinely indicated during CPR. 14 references
4/87-#9
28. DIFFERENCE IN ACID-BASE STATUS BETWEEN VENOUS AND ARTERIAL BLOOD DURING
CARDIOPULMONARY RESUSCITATION Weil, M.H., et al. N Engl J Med 315(3):153 July
17, 1986
This clinical study, from the University of Health/Science Chicago Medical
School, compared arterial and mixed venous blood gas parameters during
cardiopulmonary resuscitation in 16 patients hospitalized in an intensive or
cardiac care unit. Blood samples were obtained from five to 79 minutes after the
onset of the arrest (median, 23 minutes). The average dose of alkali given prior
to blood sampling was 130 +/- 30mEq. The average arterial pH and pCO2, values
during resuscitation (7.41 and 32mm Hg. respectively) did not differ
significantly from prearrest values, which were available in 13 patients (7.37
and 33mm Hg. respectively). In contrast, during resuscitation the average mixed
venous pH and pCO2 values were significantly altered from prearrest values (7.31
and 44mm Hg. respectively, prior to the arrest. vs. 7.15 and 74mm Hg.
respectively, during resuscitation), and were significantly different from
arterial values. Although arterial lactate levels increased significantly (p
<0.05), the bicarbonate concentrations in arterial and mixed venous blood did
not differ significantly. Overall, the venous-arterial pH gradient increased
from 0.06 +/- 0.02 prior to arrest to 0.30 +/- 0.05 during resuscitation
(p<0.001), and the pCO2, gradient increased from 11 +/- 2mm Hg to 36 +/- 6mm Hg
(p<0.001). These data confirm the findings of previous animal studies, and
suggest that respiratory acidosis occurs during cardiopulmonary resuscitation,
which may be reflected in mixed venous but not in arterial blood gases. The
findings indicate that acid-base management of cardiopulmonary arrest based on
ABGs may be inappropriate. 12 references 1/87-#10
29. ARTERIAL BLOOD GASES FAIL TO REFLECT ACID-BASE STATUS DURING CARDIOPULMONARY
RESUSCITATION: A PRELIMINARY REPORT Weil, M.H., et al, Crit Care Med 13(11):884,
November 1985
The authors, from The Chicago Medical School, analyzed arterial and venous blood
gas changes during CPR in 28 pigs. PaCO2 in the animals was maintained at
35-45torr. Determination of arterial and mixed venous blood gases before
induction of cardiac arrest and after the onset of EMD demonstrated a
significant increase in the venoarterial PCO2 and pH gradients throughout the
period of EMD (p<0.001), associated with a progressive increase in arterial
lactate levels. In surviving animals, the arterial pH increased from 7.47 at
baseline to 7.53 after four minutes of EMD, while the mixed venous pH decreased
from 7.42 to 7.26. Arterial pCO2 decreased from 38.3 torr to 22.5 torr while the
mixed venous pCO2 increased from 44.8 torr to 57.6 torr. Similar changes were
noted in nonsurvivors. After successful resuscitation, reversal of these
gradients was observed. These findings confirm observations by the authors
during previous CPR and cardiac arrest studies, as well as data obtained in
human patients, and indicate that a paradoxical acidosis in mixed venous blood
and alkalosis in arterial blood may be noted during resuscitation from cardiac
arrest. The evidence of respiratory alkalosis in arterial blood may reflect a
decrease in pulmonary blood flow and a subsequent increase in the
ventilation/perfusion ratio, while the acidosis in mixed venous blood may
reflect accumulation of CO2 proximal to the alveolar capillary bed, which more
accurately represents systemic acid-base status. These findings suggest that
during CPR administration of sodium bicarbonate based on arterial blood gas
values may be inappropriate. 4 references 5/86-#11
How do skin pigmentation, nail polish, and acrylic nails affect pulse oximetry?
The following studies show that:
1. Pulse oximetry recordings are not significantly influenced by dark skin
pigmentation in relatively stable but critically ill adults.
2. Nail polish is rarely associated with clinically significant deviations in
pulse oximeter readings, although polish removal might be helpful in some
situations.
3. Acrylic finger nails may impair the measurement of oxygen saturation,
depending on the pulse oximeter used, and may cause significant inaccuracy.
Hence, removal of artificial acrylic finger nails may assure an accurate and
precise measurement with pulse oximetry.
30. ACCURACY OF PULSE OXIMETRY IN PIGMENTED PATIENTS Bothma, P.A., et al, S Afr
Med J 86(5):594, May 1996
BACKGROUND: Prior studies of the influence of skin pigmentation on the accuracy
of pulse oximetry have yielded conflicting results. METHODS: This prospective
South African study compared the readings of three pulse oximeters (Simed S100e
[Simed Co.], Nihon Koden [Nihon Koden Corp.], and Ohmeda 3740 [Ohmeda]) with
arterial oxygen saturation by cooximetry in 100 consecutive darkly pigmented
critically ill adults. Finger probes were employed with all three pulse
oximeters, and an ear probe was also used with the Ohmeda model. RESULTS:
Arterial oxygen saturations were between 88-99% (median, 96%), with pulse
oximeter readings between 86-100%. The limits of agreement between the two
techniques were 3.4-4.5% with the Ohmeda finger probe, 3.8-5.8% with the Ohmeda
ear probe, 4.1-5.8% with the Nihon Koden, and 2.6-5.0% with the Simed, all of
which were considered to be clinically acceptable. The 95% confidence intervals
were small. The precision (standard deviation of the differences) ranged between
1.9 and 2.4%, and bias (mean of the differences) ranged between -1.0 and 1.2%.
When the arterial oxygen saturation was below 92%, there was a trend towards
lower values with pulse oximetry. Pulse oximetry results tended to be higher
than arterial oxygen saturation with the finger probes, and lower with the
Ohmeda ear probe. These differences were small and not considered clinically
significant. CONCLUSIONS: Pulse oximetry recordings were not significantly
influenced by dark skin pigmentation in these relatively stable but critically
ill adults. 25 references 1/97 - #35
31. EFFECT OF NAIL POLISH ON OXYGEN SATURATION DETERMINED BY PULSE OXIMETRY IN
CRITICALLY ILL PATIENTS Hinkelbein, J., et al, Resuscitation 72:82, January 2007
BACKGROUND: The belief that use of nail polish can be associated with misleading
results during pulse oximetry monitoring has not been extensively investigated.
METHODS: These German authors examined the effects of nine nail polish colors on
pulse oximetry readings in 50 adult Caucasian ICU patients undergoing mechanical
ventilation. A different color of polish, ranging over the entire color spectrum
from black to clear, was applied on each of nine fingernails and one fingernail
without nail polish served as a control. Pulse oximetry readings were obtained
using a SIEMENS pulse oximeter monitor SC1281 and a NELLCOR DS-100A Durasensor
finger sensor probe. Pulse oximetry readings were compared with arterial blood
gas results. RESULTS: There was a close correlation between the mean arterial
and control pulse oximetry readings (difference +0.2%). The mean difference
between the readings was greatest in the setting of black nail polish
(difference 1.6%, 95% confidence interval [CI] -4.1% to +10.6%), purple polish
(difference 1.2%, 95% CI -3.1% to +12.5%) and dark blue polish (1.1%, 95% CI
-5.7% to +9.1%), and was less than 1% with all other polish colors. For all
colors, the mean difference between arterial and pulse oximetry readings was
within the oximeter manufacturer's reported deviation of plus/minus 2% in the
oxygen saturation range of 70-100%. In general, pulse oximetry tended to
underestimate arterial oxygenation results. CONCLUSIONS: Nail polish was rarely
associated with clinically significant eviations in pulse oximeter readings,
although polish removal might be helpful in some situations. 34 references 6/07
- #34
32. ARTIFICIAL ACRYLIC FINGERNAILS MAY ALTER PULSE OXIMETRY MEASUREMENT
Hinkelbein J, Koehler H, Genzwuerker HV, Fiedler F. Resuscitation. 2007
Jul;74(1):75-82. Epub 2007 Mar 13.
INTRODUCTION: Pulse oximetry is the most common technique to monitor oxygen
saturation (SpO(2)) during intensive care therapy. However, intermittent
co-oximetry is still the "gold standard" (SaO(2)). Besides acrylic nails,
numerous other factors have been reported to interfere with pulse oximetry. Data
of measurements with artificial finger nails are not sufficiently published.
MATERIALS AND METHODS: A prospective clinical-experimental trial in mechanically
ventilated and critically ill patients of an ICU was performed. Patients were
randomly assigned to either group S (S: Siemens pulse oximeter) or group P (P:
Philips pulse oximeter) prior to the measurements. SpO(2) was determined in each
patient three times alternately in standard ((N)SpO(2)) and sideways position at
the natural nail ((N90)SpO(2)). For the reference measurements oxygen saturation
was measured by means of a haemoximeter (co-oximetry). Thereafter, SpO(2) was
obtained at the acrylic finger nail in the same way ((A)SpO(2) and (A90)SpO(2)).
Bias was calculated as DeltaS=(N)SpO(2)-SaO(2) and DeltaS=(A)SpO(2)-SaO(2).
Accuracy (mean difference) and precision (standard deviation) were used to
determine the measurement discrepancy. P<0.05 was considered significant.
RESULTS: Accuracy and precision without acrylic nails applied were comparable to
SaO(2) in both groups (n.s.). With acrylic nails applied a bias of
DeltaS=-1.1+/-3.14% for group S (P=0.00522) and a bias of DeltaS=+0.8+/-3.04%
for group P was calculated (n.s.). CONCLUSION: Acrylic finger nails may impair
the measurement of oxygen saturation depending on the pulse oximeter used and
may cause significant inaccuracy. Hence, removal of artificial acrylic finger
nails may be helpful to assure an accurate and precise measurement with pulse
oximetry.
How do abnormal hemoglobins or infusion of hydroxocobalamin affect oximetry?
Hydroxocobalamin is widely used in Europe as an alternative to other cyanide
antidotes for the treatment of patients with cyanide poisoning associated with
smoke inhalation. Increasing hydroxocobalamin concentrations are associated with
a progressive increase in the oximetric reading for carboxyhemoglobin and
methemoglobin, and a corresponding decrease in the oxygen saturation reading.
Unlike standard oximetry, the Rainbow-SET Rad-57 Pulse CO-Oximeter (Masimo) is
an eight-wavelength pulse oximeter that is capable of measuring more than two
species of human hemoglobin. At the carboxyhemoglobin and methemoglobin levels
achieved in healthy volunteers, the Rad-57 device appeared to be reliable. These
results cannot be extrapolated to the setting of critical illness or more severe
dyshemoglobinemia (or respiratory co- morbidity).
33. POTENTIAL INTERFERENCE BY HYDROXOCOBALAMIN ON COOXIMETRY HEMOGLOBIN
MEASUREMENTS DURING CYANIDE AND SMOKE INHALATION TREATMENTS Lee, J., et al, Ann
Emerg Med 49(6):802, June 2007
BACKGROUND: Hydroxocobalamin is widely used in Europe as an alternative to other
cyanide antidotes for the treatment of patients with cyanide poisoning
associated with smoke inhalation. It might be postulated, however, that the
intense light absorption produced by hydroxocobalamin can interfere with light
source-based co-oximetric blood gas measurements. METHODS: The authors, from the
University of California, Irvine, examined the effect of hydroxocobalamin
infusion (average dose, 625mg infused over 100 minutes) on oximetry readings in
rabbits. RESULTS: Increasing hydroxocobalamin concentrations were associated
with a progressive increase in the oximetric reading for carboxyhemoglobin and
methemoglobin, and a corresponding decrease in the oxygen saturation reading.
Similar patterns were observed in an in vitro study of whole blood samples with
different hydroxocobalamin concentrations. Increasing hydroxocobalamin levels
were associated with a change of up to (- )7.9% in the oxygen saturation value
and (+)14.7% in the carboxyhemoglobin value. CONCLUSIONS: The authors advise
caution in the interpretation of hemoglobin fractions measured with co-oximetry
when hydroxocobalamin is used for the treatment of cyanide poisoning and smoke
inhalation. The effects of this spectral overlap might be particularly relevant
in patients receiving continuous hydroxocobalamin infusion and when blood
sampling is performed immediately after bolus injection or "upstream" from the
hydroxocobalamin infusion site. 10 references 10/07 - #34
34. MEASUREMENT OF CARBOXYHEMOGLOBIN AND METHEMOGLOBIN BY PULSE OXIMETRY: A
HUMAN VOLUNTEER STUDY Barker, S.J., et al, Anesthesiology 105(5):892, November
2006
BACKGROUND: Standard pulse oximeters estimate arterial hemoglobin saturation via
measurement of tissue light transmission at two wavelengths. Several studies
have reported serious errors in estimation of oxygen saturation in the presence
of a dyshemoglobinemia. The Rainbow-SET Rad-57 Pulse CO-Oximeter (Masimo) is an
eight-wavelength pulse oximeter that is capable of measuring more than two
species of human hemoglobin. METHODS: The authors, from the University of
Arizona and partially funded by Masimo, examined the performance of the Rad-57
device in ten healthy volunteers exposed to carbon monoxide to produce a
carboxyhemoglobin level of 15%, and ten volunteers given an IV infusion of
sodium nitrite to induce methemoglobinemia at a level between 5-12%. Pulse
oximetry results obtained with the Rad-57 were compared with measurements
performed in arterial blood. RESULTS: At a carboxyhemoglobin level of 0-15%, the
uncertainty (a measure of test imprecision) of carboxyhemoglobin measurement
with the Rad-57 was +/- 2%. At a methemoglobin level of 0-12%, the uncertainty
with the Rad-57 was +/- 0.5%. For comparison purposes, the uncertainty of most
standard pulse oximeters for measurement of oxygen saturation at values of
70-100% is +/- 2%. CONCLUSIONS: At the carboxyhemoglobin and methemoglobin
levels achieved in these healthy volunteers, the Rad-57 device appeared to be
reliable. The authors acknowledge that these results cannot be extrapolated to
the setting of critical illness of more severe dyshemoglobinemia (or respiratory
co- morbidity). 17 references 3/07 - #34
Is room-air pulse oximetry sensitive for hypercapnia?
In this case-control study, the frequency of hypercapnia was inversely related
to the oxygen saturation on room air pulse oximetry. At an oxygen saturation
cut-off of 96%, room air pulse oximetry appears to be a sensitive screening test
for moderate hypercapnia.
35. THE SENSITIVITY OF ROOM-AIR PULSE OXIMETRY IN THE DETECTION OF HYPERCAPNIA
Witting, M.D., et al, Am J Emerg Med 23:497, July 2005
BACKGROUND: The suggestion that pulse oximetry does not reliably identify
hypercapnia has been based on studies conducted in patients receiving
supplemental oxygen. One small study reported that, at a cut-off of 96%,
room-air pulse oximetry detected all twelve patients with moderate hypercapnia.
METHODS: In this case-control study, from the University of Maryland, the charts
of 92 patients with hypercapnia and 257 patients without hypercapnia were
reviewed to assess the diagnostic utility of an oxygen saturation of 96% or
lower on room-air pulse oximetry for the detection of moderate hypercapnia
(PaCO2 above 50mm Hg). All of the patients had pulse oximetry performed on room
air followed within eight hours by measurement of arterial blood gases (this
interval was less than two hours in 43% of patients and 5-8 hours in 12%).
RESULTS: The frequency of hypercapnia was inversely related to the oxygen
saturation on room air pulse oximetry. At an oxygen saturation cut-off of 96%,
the sensitivity and specificity of room air pulse oximetry for the
identification of moderate hypercapnia were 96% and 39%, respectively, the
positive likelihood ratio (LR) was 1.6, and the negative LR was 0.1. There were
four falsely-negative cases. CONCLUSIONS: The authors acknowledge the
limitations of their study design but suggest that, at an oxygen saturation
cut-off of 96%, room air pulse oximetry appears to be a sensitive screening test
for moderate hypercapnia. 14 references 12/05 - #36
How long does it take for the equilibration of oxygen saturation using pulse
oximetry?
For patients placed on supplemental oxygen, the mean interval between initiation
of O2 and equilibration was 2.8 minutes. The mean times to equilibration in
patients with and without COPD or asthma were 3.5 and 2.5 minutes, respectively.
36. TIME TO EQUILIBRATION OF OXYGEN SATURATION USING PULSE OXIMETRY Gruber, P.,
et al, Acad Emerg Med 2(9):810, September 1995
BACKGROUND: Although it is generally believed that equilibration of the oxygen
saturation after changing the inspired oxygen concentration (FiO2) requires
20-30 minutes, the actual time required for equilibration in patients with
active medical problems has not been clearly defined. METHODS: This prospective
study, from Long Island Jewish Medical Center in New Hyde Park, NY, examined the
actual time required for equilibration of O2 saturation after changing the FiO2
in 51 adults requiring supplemental oxygen (2-4L/min by nasal cannula) in the ED
due to acute cardiac and/or pulmonary conditions. Equilibration times were
calculated on the basis of O2 saturation measurements made at one-minute
intervals for 30 minutes after a change was made in FiO2. RESULTS: For patients
placed on supplemental oxygen (43 measurements), the mean room air oxygen
saturation was 89.8%, the mean equilibration point was 95.3%, and the mean
interval between initiation of O2 and equilibration was 2.8 minutes (95%
confidence interval, 2.3-3.3 minutes). The mean times to equilibration in
patients with and without COPD or asthma were 3.5 and 2.5 minutes, respectively
(p=0.01). Upon discontinuation of supplemental oxygen (18 measurements), the
mean initial oxygen saturation was 96.6%, the mean equilibration point was
91.7%, and the mean interval to equilibration was 3.1 minutes (95% CI, 2.3-4.4
minutes), or 4.6 minutes in patients with COPD or asthma compared with 2.7
minutes in patients without these conditions. One patient in each group
demonstrated continuous variability in O2 saturation throughout the study.
CONCLUSIONS: Times to equilibration after making a change in inspired O2
concentration appear to be substantially shorter than traditionally believed.
Equilibration times in patients with COPD or asthma may be longer than in other
patients. 15 references 12/95 - #38
Should patients who are hyperventilating be given a paper bag to breathe into?
No. The author, from the University of California in San Francisco, reports
three cases in which collapse occurred in patients with myocardial ischemia or
hypoxemia who were erroneously believed to be hyperventilating and were
instructed to breathe into a paper bag. In a subsequent volunteer study, paper
bag rebreathing was associated with a substantial decrease in inspired oxygen.
37. HYPOXIC HAZARDS OF TRADITIONAL PAPER BAG REBREATHING IN HYPERVENTILATING
PATIENTS Callaham, M., Ann Emerg Med 18(6):622, June 1989
Patients who are hyperventilating are often managed with paper bag rebreathing.
The author, from the University of California in San Francisco, reports three
cases in which collapse occurred in patients with myocardial ischemia or
hypoxemia who were erroneously believed to be hyperventilating and were
instructed to breathe into a paper bag. In two of the cases, this treatment was
initiated by paramedics or by a dispatcher, and in the remaining case treatment
was initiated by a nurse. Two of the three patients sustained irreversible
cardiopulmonary arrests. In a subsequent study, 20 healthy adults
hyperventilated to an end-tidal CO2 of 20mm Hg, and were then instructed to
hyperventilate into a paper bag containing sensors for the purpose of monitoring
gas concentrations. Varying inspired concentrations of CO2 (FiCO2) were achieved
with three minutes of bag rebreathing. The FiCO2 did not reach 40mm Hg for 30%
of the subjects, and did not reach 35mm Hg for 22%, suggesting that bag
rebreathing appears to be an unreliable method of achieving an elevated FiCO2.
The inspired concentration of oxygen (FiO2) progressively declined over three
minutes of bag rebreathing, to a mean maximal decrease in O2 of 26mm Hg. The
decrease in O2 was 34mm Hg in four subjects, and 42mm Hg in one subject. The
author suggests that paper bag rebreathing may be associated with a substantial
decrease in inspired oxygen that may pose a threat to hypoxic patients. It is
further suggested that this technique not be employed unless myocardial ischemia
can be ruled out and direct measurement of the patient's oxygenation has been
performed. The author cautions against the recommendation and use of this
technique by prehospital personnel. 37 references 11/89 - #34
KEY POINTS AND RECOMMENDATIONS
1. Peripheral venous blood gases may be a reliable substitute for arterial blood
gases in the initial assessment of patients with suspected DKA.
2. Arterial blood gas analysis does not reflect the acid-base status of systemic
tissue during CPR. Respiratory alkalosis in arterial blood reflects a decrease
in pulmonary blood flow. Acidosis in mixed venous blood reflects accumulation of
CO2 proximal to the alveolar capillary bed. This mixed venous acidosis may more
accurately represent systemic acid-base status.
3. Pulse oximetry recordings are not significantly influenced by dark skin
pigmentation in stable but critically ill adults.
4. Nail polish is rarely associated with clinically significant deviations in
pulse oximeter readings, although polish removal might be helpful in some
situations.
5. Acrylic finger nails may impair the measurement of oxygen saturation
depending on the pulse oximeter used and may cause significant inaccuracy.
Hence, removal of artificial acrylic finger nails may assure an accurate and
precise measurement.
6. The Rainbow-SET Rad-57 Pulse CO-Oximeter (Masimo) is an eight-wavelength
pulse oximeter capable of measuring carboxyhemoglobin and methemoglobin levels.
7. If someone tries to put a paper bag over your head, don’t let them.
Stephen Colucciello MD
Associate Chair, Department of Emergency Medicine
Carolinas Medical Center