Ventilatory Weaning and Extubation
Tobins review of the weaning meta-analysis and MAs in general (Crit Care Med
2008;36:1)

ccm
2008 multi-center SBT protocol
Weaning from Mechanical Ventilation
Most patients fail weaning because of excessive respiratory
loads leading to muscle fatigue.
Factors that predispose to increased respiratory load
-
Increased dead space ventilation
-
Excessive CO2 production-from excessive carbohydrates in
nutritional support or disease state
-
Dynamic Hyperinflation-flattens the diaphragm and puts
respiratory musculature in less than optimal state
Phrenic nerve dysfunction-seen especially in post CABG
patients
Critical Illness Polyneuropathy
Acute necrotizing myopathy-seen in patients receiving NMBAs
and/or steroids
Metabolic abnormalities-potassium, magnesium, calcium, and
phosphate
weaning can be complicated by many
factors associated with long-term CMV that cause respiratory pump failure, such
as sedation overhang, sleep deprivation, psychological factors, malnutrition,
bronchoconstriction, and neuromuscular problems
When PaO2/FiO2>150-200 on FiO2<50% and PEEP≤8
No pressors
spont breathing
5 miutes of CPAP 5 then rapid/shallow <105-120
if pass
30-120 minutes of spont. breathing trial
Predictors of successful weaning
Clinical judgment alone has PPV of 50% and NPV of 67% (Stroetz
and Hubmayr)
Physicians usually wait too long as opposed to acting
prematurely
Frequency/VT
<100 (105)
increased time to wean in a study when added to SBTs (Crit
Care Med 2006;34:2530)
PaO2/FiO2
>200
PEEP
≤5
Cough with suctioning
Maximum Inspiratory Pressure (PIMAX)
if less than -30 predictive of successful wean if greater
than -20, predictive of unsuccessful wean
Leak Volume on Cuff Deflation
after extubation, ~5% of patients develop laryngeal edema.
Absence of a leak with cuff deflation is predictive of laryngeal edema.
Look for at least 110 cc of leak averaged over three lowest volumes.
The average duration of translaryngeal intubation was 28.1 ± 17.6 days. The
incidence of severe laryngeal edema was 36.8% (35/95). We chose 140 mL as the
threshold cuff-leak volume below which edema is indicated. The rate of cuff-leak
test positivity was 38.9% (37/95). The sensitivity and the specificity of the
test were 88.6% and 90.0%, respectively. The positive and negative predictive
values were 83.8% and 93.1%, respectively. Patients who developed severe
laryngeal edema had a smaller leak volume than those who did not, expressed in
absolute values (53.9 ± 56.2 vs. 287.9 ± 120.0 mL; p < .001) or in reltive
values (10.1 ± 10.2 vs. 55.3 ± 22.7%, p < .001). The occurrence of severe
laryngeal edema was not associated with age, gender, body weight, respiratory
failure due to pneumonia, duration of translaryngeal intubation, endotracheal
tube diameter, Acute Physiology and Chronic Health Evaluation II score, or
history of self-extubation. (Crit Care Med 34(2), February 2006, pp 409-414)
weaning parameters
- (Yang, 1991)
- a) f/Vt < 100 breaths/min./liter
- (1) calculation:
- (a) disconnect the patient from the ventilator for 1 minute
- (b) measure the spontaneous minute ventilation and the respiratory
rate
- (c) calculate the average tidal volume (in liters) as the minute
ventilation ÷ the respiratory rate
- (d) f/Vt = respiratory rate ÷ average tidal volume
- (2) in one study (Yang, 1991), this was the single best predictor of
success in weaning patients from mechanical ventilation
- (a) sensitivity 97%, specificity 64%
- b) maximal inspiratory pressure (also mistakenly referred to as "NIF")
more negative than -20
- (1) sensitivity 100%, specificity 14%
- c) minute ventilation (VE) < 10 liters/min.
- (1) sensitivity 31%, specificity 61%
- d) tidal volume (Vt) > 4 ml/kg
- (1) sensitivity 94%, specificity 39%
- e) vital capacity > 10 ml/kg
- f) FiO2 < 40% with pO2 > 60 mm Hg
Methods of Weaning
attempt to decrease respiratory load to minimum possible
identify neuromuscular dysfunction
Progressively decreased PSV or T-piece trials were equally
effective.
Once daily T-piece trials should probably be the standard,
~30 minutes with failures placed back on full support.
When evaluating IMV vs PSV vs spont breathing trials, the
latter won handily NEJM 1995;332:345-50)
Tobin's Analysis of Rapid Shallow
Bayes theorem and the performance of f/Vt (Inten Care Med 2006;32:2002)
No Condition preventing liberation
Rapid Shallow Breathing Index (Screening Test; will have false positives
but should not have false negatives)
Spont Breathing Trial
Extubate and watch 24 hours
Reintubation Following Unsuccessful Weaning
10-20% of successfully weaned patients will require
reintubation within 24 hours
Upper airway obstruction, excess secretions, respiratory
failure, CHF, encephalopathy are among the reasons
Extubation
Extubation over airway exchange cath seems to ease reintubation in patients. (Anesth
Analg 2007;105)

Steroids
another steroid for extubation study. (can j anesth 2008 / 55: 6 / pp
382–385)
for extubation: raise head 30 seconds
stick out tongue squeeze hand open eyes
Extubation: good method for extubating patient with known difficult airway
(patient may not breathe adequately after extubation; reintubation may not be
possible) over jet stylet, tube changer, or airway exchange catheter; use tube
changer to give patient 1 to 3 puffs per minute from jet ventilator while
determining if he or she able to respire; if patient not ventilating adequately,
use laryngoscope to slide new endotracheal tube into place over tube changer
Best
Anesthesiology Extubation Technique
Appropriate extubation technique facilitates the clearance of secretions
from the upper airway, avoidance of laryngospasm and maintenance of
oxygenation. A period of oxygenation prior to extubation ensures that there
is a pulmonary reservoir of oxygen to support metabolism for a few minutes
following extubation. Mouth aspiration is employed to clear secretions from
the upper airway.
When
the patient is judged fit to extubate, the pressure in the breathing circuit
is raised with manual pressure on the bag. Expansion of the chest must be
achieved to provide enough air for an effective cough. If the patient is
straining, considerable pressure may be required, is appropriate and may be
employed safely. With the chest expanded, the cuff is deflated and the tube
rapidly withdrawn while pressure is still being maintained. This maneuver
should elicit an effective cough which clears any remaining secretions from
the vocal cords.
Immediately following extubation, and before a breath of room air can be
taken, the face mask is employed to continue the administration of oxygen.
At the same time, forceful jaw thrust is provided using digital pressure
very close to the external auditory meatus. This has the benefit of
providing a clear airway in combination with a vigorous arousal stimulus.
The
only caveat when using this technique is to "Never relax for a moment!" This
technique, properly employed, is so effective that a patient may emerge,
cough and respond to commands while being aroused, and then subsequently
become stuporous, obstructed and hypoxic if left undisturbed. Trust me, I
know! Stay vigilant.
Emergency Department Extubation
TRAUMA PATIENTS CAN BE SAFELY EXTUBATED IN THE
EMERGENCY DEPARTMENT.
Scott D Weingart, Division of Emergency Critical Care, Mount
Sinai School of
Medicine, New York, NY; Jay Menaker,Hanh Truong, Kelly
Bochicchio, Thomas
M Scalea, Trauma and Critical Care, Shock Trauma Center,
Baltimore, MD
Introduction: Many trauma patients are intubated for causes that fully
resolve during their emergency department (ED) stay. If these patients could be
extubated in the ED, it would limit the need for intensive care bed admissions
and possibly allow discharge from the hospital. Hypothesis: We hypothesized that
the extubation of patients in the Emergency Department would be safe and limit
the need for hospital admission.
Methods: Data of adult trauma
patients who were intubated and then extubated in the ED during a 1 year period
at a single trauma referral center were prospectively collected for a quality
initiative. Two trained abstractors retrospectively reviewed these data and
additional information from the trauma registry and patient charts. The primary
outcome was the need for unplanned reintubation during hospitalization.
Additional outcomes were disposition and complications from the extubation.
Results: 50 eligible patients were included in the study. Reasons for the
intubation included 24 patients (48%) for combative behavior/decreased mental
status prior to CT scan, 18 patients (36%) for sedation prior to the performance
of a painful procedure, and 3 patients (6%) for seizures prior to CT scan. 100%
(95% CI 91-100%) of the patients remained extubated, with no need for unplanned
reintubation. Eight (16%) of the patients were able to be discharged from the ED
prior to admission.
Conclusions: Although our findings must be
verified in larger, controlled studies, it appears to be safe to extubate
patients in the ED, if the condition necessitating intubation has fully
resolved. This practice can reduce admission rates and limit the need for
intensive care unit beds for the patients who are admitted.
(Crit
Care Med. 2008;36(12S):46)
Post-Extubation Complications
Extubated patients are prone to aspiration because of vocal
cord dysfunction for at least 8 hours after tube removal
Consider withholding oral feeds for 24 hours or until
swallowing evaluation
Most common side effects are sore throat and hoarseness
Delayed sequelae are seen when ulcerations from the tube
heal and cause laryngeal or tracheal stenosis. High volume low pressure
cuffs have sharply decreased the incidence of stenosis
consider extubation under these conditions— awake
and responsive, adequate cough reflex, hemodynamically stable, minimal bleeding,
normothermic (not shivering), adequate renal function, and weaning parameters
met
gas ex
PaO2>60 on FiO2<0.4
Aa gradient <
PaO2/FiO2 ratio
Vital Capacity 10-15 cc/kg
NIF better than -40
Minute Vent <10 L/min
Rapid Shallow Breathing index
trials of spont breathing
can get negative pressure pulmonary edema from struggling
against closed or swollen glottis
prospective study included 76 patients with endotracheal intubation for more
than 12 h. The leak, in percent, was defined as the difference between expired
tidal volume measured just before extubation, in volume-controlled mode, with
the cuff inflated and then deflated. A gas leak around the endotracheal tube
greater than 15.5% can be used as a screening test to limit the risk of
re-intubation for laryngeal edema. (Intensive Care Med. 2002 Sep;28(9):1267-72.)
Leak test is only 80% sensitivie
Work of Breathing
Work can be defined as the force needed to effect a change. For instance, if
one were to push a chair from one side of the room to the other, the amount
of work necessary for this action could be determined by measuring how much
pressure was applied to the chair and how far the chair is moved. Thus, work
in this example would be calculated as
force x distance. WOB can be
expressed in a similar format, with force being expressed as the pressure
that is required to effect a change in volume and similarly could be
expressed as
pressure x volume, where pressure is the effort or force
generated by the respiratory muscles or ventilator, and volume is the amount
of gas exchanged as a result of the force applied.
The work load placed on
the respiratory muscles is minimal during quiet breathing, accounting for
approximately 5% of the total body oxygen consumption.[1] Even
under stress, a normal individual will be able to increase their ventilation
with only a moderate increase in total work. This is testimony to the
amazing configuration of the respiratory apparatus.
For the patient receiving mechanical ventilation, the work necessary for
effective ventilation can be visualized as either physiologic work or
imposed work (Table 1).
Table 1. Classifications of Work
| Total Work of Breathing = Physiologic + Imposed |
| Physiologic Work of Breathing = Elastic + Nonelastic |
| Imposed Work of Breathing = Ventilator Demand Valve + Circuit +
Artificial Airway |
Physiologic Work of Breathing
Much of what we know about the about the WOB originates from the early
works of Otis and coworkers.[2] They partitioned the WOB into
those forces required to overcome the elastic resistance and the frictional
resistance of the lungs, chest wall, and airways. The elastic component is
primarily influenced by inward recoil of the lungs and the outward recoil of
the chest wall. Elastic work is performed primarily during inspiration to
expand the lungs and chest wall, which creates a pressure gradient to move
gas into the lungs. Factors that contribute to the elastic WOB include the
compliance "stiffness" of the pulmonary tissue, inward recoil pressure of
the chest wall, and resistance offered by the abdominal cavity. Patients who
have an underlying disorder that results in decreased lung or chest wall
compliance will require more work to effect a given change in tidal volume.
Included in this group are those with decreased compliance secondary to
pulmonary infiltrates, fibrotic parenchymal disorders, atelectasis, and
abdominal distention.
Energy to overcome the frictional forces during quiet or active breathing
is primarily directed at overcoming flow resistance created in the airways.
This frictional component is often referred to as the nonelastic WOB.
Nonelastic work loads are generated in the airways as gas flows through them
during inspiration and expiration. Typically referred to as airways
resistance, this work load is primarily influenced by the caliber and
configuration of the airways. Acute chances in airway resistance may occur
with accumulation of airway secretions, constriction of airway smooth
muscle, edema, and inflammation of airway mucosal lining as seen during an
asthma attack. Additionally, aspiration of foreign bodies or narrowing of
the airway from extralumunal compression also increases the nonelastic WOB.
Chronic changes in airway caliber are frequently seen in patients with
chronic obstructive pulmonary disease (COPD), especially those diagnosed
with chronic bronchitis. In these patients, the airway loses some of its
natural rigidity, through repeated infections, and may partially collapse.
In normal individuals, approximately two thirds of the total WOB can be
attributed to elastic forces and the remaining one third to nonelastic
(frictional) forces (Table 2).[3] The majority of the physiologic
WOB is performed during inspiration due to the natural recoil properties of
the lung and chest wall that provide energy for exhalation. However,
patients with COPD may experience an increased WOB during exhalation. These
patients present a different challenge when considering their ventilatory
workloads. While inspiratory work loads may be normal or even reduced, it is
expiratory work that becomes elevated. Normally, the inherent elastic recoil
of lungs performs the greatest portion of expiratory work; thus, the
respiratory muscles are thought to be in a resting state during this period.
However, with a loss of elasticity, as seen in pulmonary emphysema, this
natural recoil effort is compromised and the patient is required to use
their respiratory muscles to aid in exhalation. In addition to a loss of
pulmonary elasticity, these patients often experience an increase in
resistive work during normal or forced exhalation as pleural pressure
increases, causing the airways to narrow. The viscosity of gas, especially
during mechanical ventilation, is not a major determinant in the WOB.
However, special gas mixtures containing helium have been used to
specifically reduce airways resistance.
Table 2. Physiologic Force Factors
| Elastic Work Forces |
Resistive Work Forces |
| Thoracic elastance |
Airway resistance |
| Lung elastance |
Gas viscosity |
Imposed Work of Breathing
The imposed WOB refers to the forces required from the
patient to initiate and terminate gas flow from the
ventilation (Table 3). This "imposed" work is in
addition to physiologic work and is affected by the
response of ventilator's demand valve to patient trigger
effort (sensitivity), the matching of inspiratory flow
to patient demand (synchrony), and resistance from the
artificial airway. Table 3. Imposed Work of Breathing
Factors
| 1. Response of ventilator to patient effort |
| a. Sensitivity -- trigger effort |
| b. Synchrony -- matching of ventilator
flow to patient demand |
| 2. Frictional resistance from the artificial
airway (nonelastic imposed work). |
Sensitivity refers to the effort required from the
patient to signal the ventilator for delivery of a
mandatory breath. This effort, usually referred to as
trigger effort, can either be in the form of a drop in
pressure "pressure trigger" or from a change in measured
flow "flow trigger" in the ventilator circuit. In early
ventilators, prior to microprocessor technology, the
primary method of triggering was accomplished through a
pressure signal. These ventilators, and even early
microprocessor models, had demand valves that often were
not responsive to patient effort and thus created an
increase in the WOB. With the advent of current
microprocessor technology, demand valve sensitivity is
rarely seen as a problem; however, controversy remains
over the merits of pressure triggering vs flow
triggering.
While sensitivity issues are related to patient
effort required to signal the ventilator to initiate a
breath, patient-ventilator synchrony is related to the
matching of flow from the ventilator to the patient's
inspiratory demand. At the onset of a patient-initiated
effort, the ventilator must provide flow sufficient to
meet the patient's inspiratory needs. If the flow
delivery system of the ventilator, either by design or
from the selection of a particular breath type, is not
able to meet the patient's demand, patient-ventilator
dysynchrony often occurs. Here, the patient is demanding
more output from the ventilator, at the onset of a
breath, than it can provide, and a sense of air hunger
is often experienced. The patient gets "out of synch"
with the ventilator and often tries to override the
ventilator's gas flow delivery system. This may lead to
increased effort and frustration from the patient.
Several investigators have suggested that selection of
ventilator breath types that allow for variable flow (ie,
pressure control breaths) may avoid flow dysynchrony.[4]
This benefit may be partially off-set, however, if lung
protective ventilation strategies are used.[5]
The work required to initiate and sustain gas flow
from most modern microprocessor-based ventilators is
negligible and can be adjusted through selection of
trigger modes, either pressure or flow, and selection of
breath types. However, even more critical is the imposed
WOB created by artificial airways. Care is required in
the selection of a properly sized artificial airway and
also in maintaining its patency with appropriate
bronchial hygiene procedure. Inadvertent narrowing of an
endotracheal or tracheotomy tube, from accumulated
secretions or compression, will result in a profound
increase in patient work.
|
|
Bronchial hygiene procedures that are frequently used to minimize the WOB
include airway suctioning, aerosol treatments, and secretion mobilization
with chest physical therapy or percussive breathing maneuvers. These
therapies are directed at improving lung compliance and reducing airway
resistance. Positioning the patients to optimize respiratory muscle function
can also be used to reduce the WOB.[8]
Pressure Support Ventilation
PSV is a patient assist maneuver that is applied to
spontaneous breaths during mechanical ventilation.
Pressure support assists the patient's effort through a
preset pressure level applied during inspiration. At the
onset of a spontaneous breath, the ventilator delivers
flow to the patient sufficient to achieve the prescribed
pressure level. As airway pressure is maintained, flow
gradually decelerates in response to the patient's
decreasing inspiratory demand, and then at a
predetermined point of minimal flow, spontaneous
inspiration is terminated and expiration begins. By
assisting the patient's inspiratory effort with a preset
level of pressure, PSV can be used to overcome both
elastic and nonelastic WOB. Lower levels of PSV are
envisioned as assisting the patient in flow-resistive
work (nonelastic) associated with the imposed WOB from
the artificial airway. As the level of PSV is increased,
not only is the nonelastic work load decreased but
elastic work load is also reduced. [9]
Although some practitioners have advocated using a level
of PSV that performs all of the WOB, PSV primarily is
used to overcome artificial airway resistance and to
partially assist the patient's elastic work. Adjusting
the level of PSV is a topic of continued interest, even
after 15 plus years of clinical use. Many clinicians
initiate PSV using a predetermined minimal setting (eg,
5 cm H 2O) while others adjust PSV to achieve
a minimal spontaneous tidal volume based on ideal body
weight (eg, 5-8 mL/kg ideal body weight). Establishing a
spontaneous tidal volume with PSV is partially based on
the work of Otis and coworkers. [2] Based on
the patient's underlying lung pathology, either primary
restrictive or obstructive, a least WOB point is
determined. For obstructive patients, this would
generally equate to spontaneous tidal volumes of
approximately 7-8 mL/kg ideal body weight, while
restrictive patients have least WOB points at tidal
volumes closer to 4-5 mL/kg ideal body weight. In either
situation, the goal is to assist the patient in
generating a tidal volume and thus adopting a
concomitant spontaneous rate that results in a desirable
level of ventilatory work.
- Pressure support adjustment: adjust PSV to
achieve the desired tidal volume-breathing frequency
combination based on the patient's least work of
breathing point.
In older-generation ventilators, the rise to the
preset PSV pressure and the level of flow that
terminated each inspiration was preset and usually not
adjustable. As will be discussed in the next sections,
these criteria can now be adjusted to fit individual
patient needs.
|
|
Rise Time
Rise time (RT) refers to the time required by the
ventilator to achieve the preset pressure during a
pressure breath including PSV or pressure control
ventilation (PCV). RT can be envisioned as a means to
adjust the attack rate at which the ventilator
establishes a preset pressure (Figure 2) and indirectly
control peak flow. Several investigators have suggested
using RT to adjust flow delivery to establish a more
synchronous interaction between ventilator and patient.
By establishing a peak flow that is synchronous with the
patient's ventilatory pattern and underlying lung
disease, WOB can be optimized. [10] Take, for
instance, a patient with COPD who has significant airway
remolding with increased airway resistance (increased
nonelastic work). A breathing pattern that favors lower
initial flow rates may result in less flow turbulence
and improved gas distribution to those lung units distal
to partially obstructed airways. In this situation, a
lower RT may be preferred. For those patients with
restrictive disorders, a breathing pattern of smaller
tidal volumes and faster rates often results in the
least amount of energy expenditure during spontaneous
ventilation as previously discussed. This pattern
necessitates that the ventilator provide fast initial
peak flows to satisfy the patient's demand. Setting a
faster RT for these patients may lead to improved
patient-ventilator synchrony and reduce WOB levels. [11]
Figure 2. Changes in peak flow at various
rise time settings.
|
|
Expiratory Sensitivity -- Flow Termination
The cycling from inspiration to expiration on a
spontaneous breath is activated once flow has
decelerated to percent of initial peak flow. Prior to
the current generation of mechanical ventilators, the
flow termination point was preset by many manufacturers
at 25% of peak flow. Thus, if the initial peak flow on a
spontaneous breath was 100 LPM, inspiration would
proceed until flow from the ventilator decelerated to
25% of its initial level, or in this example 25 LPM
(Figure 3). The deceleration in flow, from the
ventilator, is in direct response to patient demand. As
patient demands decrease, flow decelerates, and at a
given point inspiration is terminated. In a perfect
situation, the termination of inspiration would coincide
with patient effort. Having an exact match between the
ventilator and patient's neural drive would avoid under
or over shooting flow delivery and result in a
synchronous transition between inspiration and
expiration.
Figure 3. Expiratory sensitivity (Esens).
While today's ventilators lack the sophistication to
exactly match flow termination with the patient's neural
respiratory drive, many of them do allow for the setting
of a flow termination point using expiratory sensitivity
(Esens). This feature provides a means to adjust the
termination of spontaneous inspiration based on a
percent of peak flow. Again, assume that a patient has a
peak spontaneous flow of 100 LPM (Figure 3). At an Esens
of 25%, inspiration would last until flow decelerated to
25 LPM, which in this example resulted in a spontaneous
inspiratory time of 0.60 seconds (used for descriptive
purposes only). If the Esens was set to 50%, and the
patient again has an initial peak flow of 100 LPM, the
breath would terminate when flow decelerated to 50% or
50 LPM, which would decrease spontaneous inspiratory
time to 0.30 seconds. Or if a longer spontaneous
inspiratory time is desirable, then Esens would be set
at 10% and inspiration would continue until flow has
decelerated down to 10 LPM (10% of 100 LPM), resulting
in a longer spontaneous inspiratory time. In each of
these hypothetical situations, spontaneous inspiratory
time was selectively altered by adjusting Esens.
Selective use of various flow termination points has
been studied by several investigators in patients with
COPD.[12] Imagine a patient with emphysema
who is on PSV and CPAP. Due to the nature of the
patient's airway disease, expiratory flows are
diminished and at times may never decrease to the set
Esens level. The patient may have to forcefully end
spontaneous inspiration by contracting their expiratory
muscles, to increase the signal to the ventilator to end
spontaneous inspiration, which may lead to further air
trapping and autoPEEP. Several strategies could correct
this problem, including a resetting of the flow
termination point by increasing the Esens level. For
those patients with restrictive lung disorders, a
lengthening of spontaneous inspiratory time may be
desirable to increase tidal volume at any given level of
pressure support. Here, a decrease in Esens may prove
useful.[13]
Adjusting PSV, RT, or Esens is at the present time
not an exact science. Consideration of the patient's
underlying cardiopulmonary disease, any acute situations
that may influence the neurologic drive for ventilation,
and conditions that alter or impair respiratory muscle
function must all be taken into account when determining
the most appropriate setting for maneuvers. Monitoring
spontaneous breathing frequency and tidal volume may be
useful in determining if the rise in pressure matches
the elastic and nonelastic properties of a patient's
pulmonary system. At the "optimal" setting for these
strategies, spontaneous tidal volume should be in normal
range (5-8 mL/kg ideal body weight), with obstructive
patients favoring the higher end of this range and
conversely restrictive patients in the lower end of this
range. Tidal breathing in this range should result in a
spontaneous breathing frequency that results in the
lowest WOB for each patient type (Table 5). However,
care should be taken when adjusting RT or Esens to
assess patient-ventilator synchrony. This may be
accomplished by observing the retraction of the
inspiratory muscles, in the case of RT, or expiratory
muscles, when using Esens, to ensure that the patient is
not "fighting" the ventilator. Using the advanced
graphic features on modern ventilators affords
clinicians the opportunity to observe patient-ventilator
synchrony.[10]
Table 5. Obstructive vs Restrictive
| Patient Type |
Pressure Support |
Rise Time |
Esens (flow termination) |
| Obstructive |
7-8 mL/kg IBW tidal volume |
Lower setting to reduce turbulence |
Higher setting to avoid air trapping |
| Restrictive |
4-5 mL/kg IBW tidal volume |
Higher setting to match increased drive |
Lower setting to increase VT |
Esens = expiratory sensitivity; IBW = ideal body
weight; VT = tidal volume
|
|
Automatic Tube Compensation
As previously mentioned, PSV allows for a preset level
of pressure to be applied during each spontaneous
breath. Critical to using PSV is determining the
appropriate level of pressure that is set by the
operator. Pressure settings may range from minimal
settings sufficient to provide spontaneous tidal volumes
based on a least WOB point or may be set such that the
level of PSV provides tidal volumes similar to those
used for mandatory breaths. Regardless of the initial
level of PSV, the goal is to reduce PSV to a minimal
setting prior to ventilator discontinuance. A minimal
setting for PSV is often defined as that level of
support that is sufficient in overcoming only imposed
WOB from the artificial airway, while encouraging the
patient to perform the remaining WOB (elastic work).
Determining a minimal PSV level that accomplishes this
has challenged practitioners since the inception of PSV,
and there is often confusion on what pressure should be
used to achieve these outcomes. Using too high of a PSV
level may mask a problem and the patient could develop
respiratory muscle fatigue once the ventilator is
discontinued. On the other hand, too low of a "minimal"
setting may result in the patient exhibiting an
undesirable ventilator pattern while on the ventilator,
and discontinuance from the ventilator may be delayed.
Automatic tube compensation (ATC), or what is sometimes
referred to as tube compensation (TC), provides a
partial solution to this dilemma. TC allows the
ventilator to autoregulate the PSV level based on the
resistance of the artificial airway. Pressure support is
adjusted, by ventilator, throughout each spontaneous
inspiratory period in response to changes in airway
resistance. While the mechanism for accomplishing this
differs between various ventilator manufacturers, the
basic premise is that once the fixed diameter and an
approximation of the tube length is known, the
ventilator can determine artificial airways resistance.
Based on a now known endotracheal or tracheotomy tube
resistance factor, the ventilator calculates a
"targeted" carinal pressure, with carinal pressure being
pressure at the distal end of the artificial airway. TC
attempts to minimize the difference between the proximal
and carinal pressures, thus "erasing" the presence of
the imposed WOB from the artificial airway. At the onset
of the spontaneous breath, flow is delivered from the
ventilator through the airway and PSV is applied to
create a constant carinal pressure. Pressure is then
regulated proportional to the inspiratory flow during
each spontaneous breath. This "automatic" titration of
PSV occurs within minimum and maximum limits set on the
ventilator to avoid pressure "runaways". As with
traditional PSV, the breath can be tailored using RT and
Esens to match flow delivery and inspiratory time to the
patient's underlying lung disease. The efficiency of ATC
has been demonstrated by several investigators.[14]
While these studies show a dramatic decrease in the
imposed WOB with ATC, factors such as kinking of the
tube and accumulation of secretion within the airway may
reduce the effectiveness of this maneuver.[15]
- Tube compensation takes the guesswork out of
setting a "minimal" level of pressure support.
TC may provide another method to assess certain
patients for extubation. Patients who would typically be
placed on a set level of "minimal" PSV or removed from
the ventilator and placed on supplementary oxygen may
benefit from ATC. In this situation, ATC takes the guess
work out of selecting a "minimal level of PSV" and
negates the need to take the patient off of the
ventilator so they may be assessed for extubation. In a
sense, ATC can be seen as a form of "electronic
extubation" providing the practitioner with a glimpse of
how the patient will respond if extubated, while
avoiding the hazards and complications of premature
extubation and the need for immediate reintubation.
|
|
Steroids for poor leak test folks
Solumedrol 40 mg IV x 1 followed by extubation in 24 hours
or 40 mg IV Q6 with extubation 24 hours later
(Crit Care Med 2006;34(5):1345)
Bailey Maneuver
Insert LMA before removing the tube

Only two components of a semiquantitative assessment of the need for airway
care were associated with successful extubation: spontaneous cough (P =
0.01) and suctioning frequency (P = 0.001).
Steroids for prevention of postextubation edema (Lancet
2007'369:1083)
start solumedrol 20 mg 12 hrs before extubation and every
four hours thereafter until tube removal
Dexamethasone to prevent postextubation airway
obstruction in adults: a prospective, randomized, double-blind,
placebo-controlled study
Chao-Hsien Lee
, Ming-Jen Peng
and Chien-Liang
Wu

Critical Care 2007, 11:R72 doi:10.1186/cc5957
Abstract (provisional)
The complete article is available as a
provisional PDF. The
fully formatted PDF and HTML versions are in production.
Introduction
Prophylactic steroid therapy to reduce the occurrence of postextubation
laryngeal edema is controversial. Only a limited number of prospective trials
involve adults in an intensive care unit. The purpose of this study was to
ascertain whether administration of multiple doses of dexamethasone to
critically ill, intubated patients reduces or prevents the occurrence of
postextubation airway obstruction. Another specific objective of our study was
to investigate whether an after-effect (i.e., a transient lingering benefit)
exists 24 hours after the discontinuation of dexamethasone.
Methods
A randomized, placebo-controlled, double-blind trial was conducted in an
adult medical intensive care unit of a tertiary care hospital. 86 patients who
had been intubated for more than 48 hours with a cuff leak volume < 110 mL and
met weaning criteria were randomly assigned to receive either dexamethasone (5
mg; n = 43) or placebo (normal saline; n = 43) every 6 hours for a total of four
doses on the day preceding extubation. Cuff leak volume was measured before the
first injection, one hour after each injection and 24 hours after the 4th
injection. Extubation was carried out 24 hours after the last injection.
Postextubation obstruction (defined as the presence of stridor) was recorded
within 48 hours of extubation.
Results
Administration of dexamethasone during the 24-hour period preceding
extubation resulted in a statistically significant increase in the cuff leak
volume (p < 0.05). The significant increase of CLV and change of CLV relative to
baseline tidal volume (%) was not only throughout the treatment period, but also
24 hours after the last dexamethasone injection. The incidence of postextubation
stridor was significantly lower in the dexamethasone group than in placebo group
(10% [4/40] versus 27.5% [11/40], p =0.037), whereas there was no significant
difference in reintubation rate between the two groups (2.5% [1/40] versus 5%
[2/40], p =0.561).
Conclusions
Prophylactic administration of multiple-dose dexamethasone is effective in
reducing the incidence of postextubation stridor in the adult patients at high
risk for postextubation laryngeal edema. The after-effect of dexamethasone may
validate the reduced incidence of postextubation stridor in delay of extubation
within 24 hours after multiple doses dexamethasone. Trial Registration:
NCT00452062
(Chest. 2007; 131:1742-1746)
Postobstructive Pulmonary Edema A Case for Hydrostatic Mechanisms
Measurement of the edema fluid/plasma protein ratio and the
presence of net alveolar fluid clearance in 10 patients with postobstructive
pulmonary edema supports a hydrostatic mechanism for edema fluid formation. The
predominantly fast rates of alveolar fluid clearance may explain the rapid
resolution of clinical postobstructive pulmonary edema that is typically
described.
Table 63-17 -- Criteria for operating room or
postanesthesia care unit extubation of trauma patients (from Dutton)
| Mental Status |
| Resolution of intoxication |
| Able to follow commands |
| Noncombative |
| Pain adequately controlled |
| Airway Anatomy and Reflexes |
| Appropriate cough and gag |
| Ability to protect airway from aspiration |
| No excessive airway edema or instability |
| Respiratory Mechanics |
| Adequate tidal volume and respiratory rate |
| Normal motor strength |
| Required Fio2 < 0.50 |
| Systemic Stability |
| Adequately resuscitated (see above) |
| Small likelihood of urgent return to the operating room |
| Normothermic, without signs of sepsis |
Tue Jan 6, 2009 10:03 am (PST)
Regular lidociane in the ETT cuff will not work nearly as well.
Adding sodium bicarbonate changes the PKA and allows more diffusion
through the PVC cuff membrane. Estebe showed for the same dose with
regular lidocaine in the ETT cuff only 1% of the neutral base will
diffuse at 6 hours versus 60% when lidocaine and bicarbonate was
added. I have measured the cuff pressures with an arterial line
transducer and the pressure volume compliance curve is slightly
steeper with fluid versus air but not significantly. I add 3cc of 2%
lidocaine then add 1cc increments of bicarbonate until I have an
occlusive seal. You have a slightly less volume at the end of the
case and occasionally you may need to add 1cc during the case but not
often. Narcotic at the end allows even better tolerance of the ETT
on
emergence .This works well for cases greater than 1 hour. For
short
cases LTA lidocaine at the beginning helps
Gary Yurina
Post-extubation stridor review
Home |
Disclaimer |
Contact |
Podcast